TRANSONIC SYSTEMS, INC. FDA Inspection Record

FEI 1319030 · Inspections, Form 483s, warning letters and import refusals

TRANSONIC SYSTEMS, INC. is an FDA-registered facility (FEI 1319030) located in ITHACA, NY, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Feb 1, 2018: No Action Indicated (NAI)
Inspections on record
5 (1 OAI, 4 NAI)
Last OAI inspection
Mar 21, 2013
Warning letters
1 listed by FDA for this FEI, latest Jun 19, 2013
Recalls as the recalling firm
4, latest Feb 15, 2018

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Feb 1, 2018No Action Indicated (NAI)Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Supplier oversight (21 CFR 820.50(a)(2))
Jun 16, 2017No Action Indicated (NAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Complaints (21 CFR 820.198(a))
Design validation - user needs and intended uses (21 CFR 820.30(g))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
and 1 more
Nov 19, 2013No Action Indicated (NAI)Documentation (21 CFR 820.100(b))
Mar 21, 2013Official Action Indicated (OAI)No citations listed
Jan 26, 2010No Action Indicated (NAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Document change records, maintained. (21 CFR 820.40(b))
Results of investigation (21 CFR 820.198(e)(6))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 15, 2018Class IIZ-1776-2018Transonic ADT1018 Flow QC Clear Advantage Tubing, Catalog No. ADT1018, ADT1018-40 The…
The sterility of the device cannot be assured. There is a potential for an increased risk of infection.
Jul 27, 2015Class IIZ-2720-2015Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow… and 9 more products
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when…
Feb 17, 2015Class IIZ-1295-2015Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis…
Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user…
Mar 28, 2013Class IIZ-1147-2013HCS3011 AV Loop Kit For use with Transonic CO status System Only, -HCS3021 AV Loop Kit For…
Transonic Systems Inc. have received complaints of blood leakage involving the AV Loop Kit.

FDA record score

1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

6 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1776-2018, this FEI as the recalling firmFeb 15, 20184 x 0.251
Class II recall Z-2720-2015, this FEI as the recalling firmJul 27, 201510 years or older, so it adds no points0
Class II recall Z-1295-2015, this FEI as the recalling firmFeb 17, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJun 19, 201310 years or older, so it adds no points0
Class II recall Z-1147-2013, this FEI as the recalling firmMar 28, 201310 years or older, so it adds no points0
Inspection classified OAIMar 21, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing TRANSONIC SYSTEMS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check TRANSONIC SYSTEMS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Transonic Systems Inc
Location
ITHACA, NY (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.