TRANSONIC SYSTEMS, INC. FDA Inspection Record
TRANSONIC SYSTEMS, INC. is an FDA-registered facility (FEI 1319030) located in ITHACA, NY, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Feb 1, 2018: No Action Indicated (NAI)
- Inspections on record
- 5 (1 OAI, 4 NAI)
- Last OAI inspection
- Mar 21, 2013
- Warning letters
- 1 listed by FDA for this FEI, latest Jun 19, 2013
- Recalls as the recalling firm
- 4, latest Feb 15, 2018
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 1, 2018 | No Action Indicated (NAI) | Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Supplier oversight (21 CFR 820.50(a)(2)) |
| Jun 16, 2017 | No Action Indicated (NAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Complaints (21 CFR 820.198(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) and 1 more |
| Nov 19, 2013 | No Action Indicated (NAI) | Documentation (21 CFR 820.100(b)) |
| Mar 21, 2013 | Official Action Indicated (OAI) | No citations listed |
| Jan 26, 2010 | No Action Indicated (NAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Document change records, maintained. (21 CFR 820.40(b)) Results of investigation (21 CFR 820.198(e)(6)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 15, 2018 | Class II | Z-1776-2018 | Transonic ADT1018 Flow QC Clear Advantage Tubing, Catalog No. ADT1018, ADT1018-40
The… The sterility of the device cannot be assured. There is a potential for an increased risk of infection. |
| Jul 27, 2015 | Class II | Z-2720-2015 | Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC
Product Usage:
to measure flow… and 9 more products Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when… |
| Feb 17, 2015 | Class II | Z-1295-2015 | Transonic Hemodialysis Monitor, Model HD03 --
Product Usage: The Hemodialysis… Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user… |
| Mar 28, 2013 | Class II | Z-1147-2013 | HCS3011 AV Loop Kit For use with Transonic CO status System Only, -HCS3021 AV Loop Kit For… Transonic Systems Inc. have received complaints of blood leakage involving the AV Loop Kit. |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1776-2018, this FEI as the recalling firm | Feb 15, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2720-2015, this FEI as the recalling firm | Jul 27, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1295-2015, this FEI as the recalling firm | Feb 17, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jun 19, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1147-2013, this FEI as the recalling firm | Mar 28, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Mar 21, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing TRANSONIC SYSTEMS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check TRANSONIC SYSTEMS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Transonic Systems Inc
- Location
- ITHACA, NY (US)
- Source
- device