MEDELA LLC FDA Inspection Record
MEDELA LLC is an FDA-registered facility (FEI 1419937) located in MCHENRY, IL, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Aug 27, 2014: No Action Indicated (NAI)
- Inspections on record
- 8 (2 OAI, 2 VAI, 4 NAI)
- Last OAI inspection
- May 25, 2012
- Warning letters
- 1 listed by FDA for this FEI, latest Feb 25, 2011
- Recalls as the recalling firm
- 4, latest Jun 5, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 27, 2014 | No Action Indicated (NAI) | No citations listed |
| Feb 26, 2014 | No Action Indicated (NAI) | No citations listed |
| May 25, 2012 | Official Action Indicated (OAI) | Food-contact - unlawful indirect additives (21 CFR 110.40(a)) Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 3 more |
| Nov 21, 2011 | No Action Indicated (NAI) | No citations listed |
| Aug 9, 2011 | Voluntary Action Indicated (VAI) | No citations listed |
| Nov 1, 2010 | Official Action Indicated (OAI) | No citations listed |
| Jan 21, 2010 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Investigation of device failures (21 CFR 820.198(c)) Uniform and timely processing (21 CFR 820.198(a)(1)) Design verification - documentation (21 CFR 820.30(f)) and 3 more |
| Apr 2, 2009 | No Action Indicated (NAI) | Design control - no procedures (21 CFR 820.30(a)) Essential design outputs (21 CFR 820.30(d)) Documentation (21 CFR 820.100(b)) Design validation - documentation (21 CFR 820.30(g)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 5, 2026 | Class II | Z-2645-2026 | Brand Name: Medela
Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN
Model/Catalog… May contain incorrect connectors and will not attach or misconnect to ENFit connections and is thus incompatible. Issue could lead to slight… |
| Dec 2, 2019 | Class II | Z-0773-2020 | Invia Foam Dressing Kit with FitPad - X-Large (3 pcs) SKU # 0876224.
The Invia Foam… Invia Foam Dressing Kit with FitPad could have holes in the sterile packaging. |
| Dec 5, 2018 | Class III | Z-2061-2019 | Breast Milk Transfer Lid for use with Oral Syringe Connectors
Model: 90003S-100 The sterility may be compromised due to a potential breach in the packaging, even though it may appear intact. |
| Feb 7, 2018 | Class II | Z-2528-2018 | Invia Motion Negative Pressure Wound Therapy System. Labeled as:
a. Endure Negative… Device may display a battery missing error. |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2645-2026, this FEI as the recalling firm | Jun 5, 2026 | 4 x 1 | 4 |
| Class II recall Z-0773-2020, this FEI as the recalling firm | Dec 2, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2528-2018, this FEI as the recalling firm | Feb 7, 2018 | 4 x 0.25 | 1 |
| Inspection classified OAI | May 25, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 9, 2011 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Feb 25, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Nov 1, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 21, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing MEDELA LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check MEDELA LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Medela Inc
- Location
- MCHENRY, IL (US)
- Source
- device