Northgate Technologies Inc. FDA Inspection Record

FEI 1450997 · Inspections, Form 483s, warning letters and import refusals

Northgate Technologies Inc. is an FDA-registered facility (FEI 1450997) located in Elgin, IL, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Apr 19, 2022: Voluntary Action Indicated (VAI)
Inspections on record
4 (3 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
6, latest Apr 13, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 19, 2022Voluntary Action Indicated (VAI)Servicing - Lack of or inadequate procedures (21 CFR 820.200(a))
Jun 1, 2017No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Investigation of device failures (21 CFR 820.198(c))
Apr 3, 2012Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Analyzing service report (21 CFR 820.200(b))
Nov 13, 2009Voluntary Action Indicated (VAI)Incoming inspection, testing, verification (21 CFR 820.80(b))
Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 13, 2022Class IIZ-1235-2022AUTOLITH Touch 1.9F 375cm Biliary Electrohydraulic Lithotripter Probe
Product labeled incorrectly.
Oct 25, 2021Class IIZ-0333-2022Boston Scientific AUTOLITH TOUCH Bipolar Electrohydraulic Lithotripter, REF…
The languages for Russian, Romanian, Slovak, and Czech have the power settings for high and low switched in the device.
Jan 15, 2021Class IIZ-1147-2021ConMed 50L abdominal insufflator, Catalog Number 72-00258-0 / GS2000 - Product Usage: The… and 1 more product
An internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.
Mar 10, 2020Class IIZ-1883-2020Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the… and 2 more products
As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.
May 24, 2017Class IIZ-0073-2018Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model…
The inner and outer labels on the product contain incorrect reference number of 9-202-3751DS. According to Northgate Technologies, the…
Nov 21, 2016Class IIZ-0866-2017Insufflator, laparoscopic The ConMed GS2000, Smith & Nephew 500, Novadaq 50L and…
Device could cause a patient overpressure situation without any visual or auditory indication or warning

FDA record score

10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIApr 19, 20226 x 0.63.6
Class II recall Z-1235-2022, this FEI as the recalling firmApr 13, 20224 x 0.62.4
Class II recall Z-0333-2022, this FEI as the recalling firmOct 25, 20214 x 0.251
Class II recall Z-1147-2021, this FEI as the recalling firmJan 15, 20214 x 0.251
Class II recall Z-1883-2020, this FEI as the recalling firmMar 10, 20204 x 0.251
Class II recall Z-0073-2018, this FEI as the recalling firmMay 24, 20174 x 0.251
Class II recall Z-0866-2017, this FEI as the recalling firmNov 21, 201610 years or older, so it adds no points0
Inspection classified VAIApr 3, 201210 years or older, so it adds no points0
Inspection classified VAINov 13, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Northgate Technologies Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Northgate Technologies Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Northgate Technologies, Inc.
Location
Elgin, IL (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.