Gentherm Medical, LLC FDA Inspection Record
Gentherm Medical, LLC is an FDA-registered facility (FEI 1516825) located in Cincinnati, OH, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Oct 9, 2020: No Action Indicated (NAI)
- Inspections on record
- 7 (1 OAI, 2 VAI, 4 NAI)
- Last OAI inspection
- Dec 14, 2009
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 13, 2010. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 15, latest Jul 13, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 9, 2020 | No Action Indicated (NAI) | No citations listed |
| Aug 30, 2018 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate procedures (21 CFR 820.100(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) |
| Jan 18, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 2 more |
| Mar 1, 2013 | No Action Indicated (NAI) | No citations listed |
| Nov 18, 2011 | No Action Indicated (NAI) | Investigation of device failures (21 CFR 820.198(c)) |
| Dec 14, 2009 | Official Action Indicated (OAI) | No citations listed |
| Jul 15, 2009 | Voluntary Action Indicated (VAI) | No citations listed |
Warning letters
- Jul 13, 2010: Warning letter listed on FDA's Data Dashboard. View compliance record (93004).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 13, 2026 | Class II | Z-2968-2026 | Blanketrol II Model 222S; Thermal regulating system. and 2 more products Overheating of the 115V CPU board may cause the device to shut down. |
| Nov 22, 2021 | Class II | Z-0496-2022 | Electri-Cool II Portable Cold Therapy Unit, Model Number 767, Part Number 86193 This recall has been initiated to provide a labeling update with revised instruction how to operate the unit with Cold Therapy pads. |
| Jun 17, 2021 | Class II | Z-2092-2021 | Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used… There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms… |
| Jul 19, 2019 | Class III | Z-2291-2019 | MICRO-TEMP LT System, Model 749- Localized Heat Therapy System is intended to warm a… Updated Manual-Warnigs have been added stating that exceeding 40'C for extended periods may cause tissue damage |
| Jul 2, 2019 | Class II | Z-0417-2020 | Blanketrol III Hyper-Hypothermia System- body hypothermia system temperature management… and 2 more products Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added… |
| Jun 20, 2019 | Class II | Z-2355-2019 | Norm-O-Temp
model 111Z and 1 more product The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage. |
| Dec 19, 2018 | Class II | Z-0825-2019 | Hemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE
Product Usage:
The… Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched… |
| Sep 4, 2018 | Class II | Z-0063-2019 | Hemotherm 400CE, Catalog Numbers 86022 & 86023
The Hemotherm¿ Dual Reservoir… Potential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended… |
| Nov 27, 2017 | Class II | Z-0629-2018 | Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR
Product Usage:
A… Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated… |
| Mar 9, 2017 | Class II | Z-1825-2017 | Hemotherm CE, model 400CE, Heater-Cooling Device Improperly rated fuses were used in the manufacture of the device. |
FDA record score
14 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
18 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2968-2026, this FEI as the recalling firm | Jul 13, 2026 | 4 x 1 | 4 |
| Inspection classified VAI | Jan 18, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0496-2022, this FEI as the recalling firm | Nov 22, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2092-2021, this FEI as the recalling firm | Jun 17, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0417-2020, this FEI as the recalling firm | Jul 2, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2355-2019, this FEI as the recalling firm | Jun 20, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0825-2019, this FEI as the recalling firm | Dec 19, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0063-2019, this FEI as the recalling firm | Sep 4, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0629-2018, this FEI as the recalling firm | Nov 27, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1825-2017, this FEI as the recalling firm | Mar 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0654-2017, this FEI as the recalling firm | Nov 7, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1129-2015, this FEI as the recalling firm | Jan 7, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1679-2014, this FEI as the recalling firm | Jan 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2209-2013, this FEI as the recalling firm | Sep 3, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0443-2014, this FEI as the recalling firm | May 15, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jul 13, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Dec 14, 2009 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 15, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Gentherm Medical, LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Gentherm Medical, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Gentherm Medical, LLC → Check this supplier
Facility details
- Company
- Gentherm Medical, LLC
- Location
- Cincinnati, OH (US)
- Source
- device