National Biological Corp FDA Inspection Record
National Biological Corp is an FDA-inspected facility (FEI 1521608) located in Beachwood, OH, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Aug 18, 2021: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (2 OAI, 3 VAI, 1 NAI)
- Last OAI inspection
- Feb 14, 2019
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jul 11, 2017. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Jan 27, 2017
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 18, 2021 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Nonconforming product control (21 CFR 820.90(a)) Investigation of device failures (21 CFR 820.198(c)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) and 3 more |
| Mar 13, 2020 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b)) Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) and 1 more |
| Feb 14, 2019 | Official Action Indicated (OAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) and 1 more |
| Mar 20, 2017 | Official Action Indicated (OAI) | Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3)) Nonconforming product control (21 CFR 820.90(a)) Review and evaluation for investigation (21 CFR 820.198(b)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) and 5 more |
| Apr 4, 2013 | Voluntary Action Indicated (VAI) | Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Nonconforming product control (21 CFR 820.90(a)) and 2 more |
| Nov 13, 2009 | No Action Indicated (NAI) | DHR content (21 CFR 820.184) Lack of System for Determining MDR Events (21 CFR 803.17(a)(2)) |
Warning letters
- Jul 11, 2017: Warning letter listed on FDA's Data Dashboard. View compliance record (523732).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 27, 2017 | Class II | Z-1683-2017 | Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a… The wiring may allow the lamps to turn on with the key switch rather than the timer. |
| Apr 9, 2013 | Class III | Z-1778-2013 | Dermalight 80 Model Number UVB-080, one unit per box.
The Dermalight 80 is a… The incorrect manual may have been packaged with this product. UVB-080 devices may have included the manual for a UVA-083 device. |
| Apr 8, 2013 | Class III | Z-1779-2013 | Dermalight 80 Model Number UVB-083, one unit per box.
The Dermalight 80 is a… The incorrect manual may have been packaged with this product. UVB-083 devices may have included the manual for a UVA-080 device. |
FDA record score
27 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified OAI | Feb 14, 2019 | 30 x 0.25 | 7.5 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jul 11, 2017 | 30 x 0.25 | 7.5 |
| Inspection classified OAI | Mar 20, 2017 | 30 x 0.25 | 7.5 |
| Inspection classified VAI | Aug 18, 2021 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Mar 13, 2020 | 6 x 0.25 | 1.5 |
| Class II recall Z-1683-2017, this FEI as the recalling firm | Jan 27, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Apr 4, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing National Biological Corp as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for National Biological Corp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about National Biological Corp → Check this supplier
Facility details
- Company
- National Biological Corp
- Location
- Beachwood, OH (US)
- Source
- inspections