RADIOMETER AMERICA, INC. FDA Inspection Record

FEI 1523456 · Inspections, Form 483s, warning letters and import refusals

RADIOMETER AMERICA, INC. is an FDA-registered facility (FEI 1523456) located in Middleburg Heights, OH, US. RegDataLab tracks 2 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Mar 2, 2015: No Action Indicated (NAI)
Inspections on record
2 (1 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
11, latest Dec 8, 2014

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 2, 2015No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Justification for not reporting (21 CFR 806.20(b)(4))
Aug 18, 2011Voluntary Action Indicated (VAI)Rationale documented for no investigation (21 CFR 820.198(b))
Servicing - Lack of or inadequate procedures (21 CFR 820.200(a))
Supplier oversight (21 CFR 820.50(a)(2))

Recalls

Showing the 10 most recent of 11 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 8, 2014Class IIZ-1046-2015ABL90 FLEX analyzer; Model number 393-090. A portable, automated analyzer that measures…
The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a…
Oct 29, 2014Class IIZ-1293-2015Crea A and B membranes, packaged 4 per box Used in conjunction with the ABL800 FLEX…
Negative drift on QC during in-use lifetime of CREA membranes can be observed.
Oct 20, 2014Class IIZ-0873-2015HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for…
Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf…
Jul 28, 2014Class IIZ-2597-2014RADIANCE v. 2.4X, 2.5X and 2.6X Product Usage: RADIANCE is a data processing…
Oct 15, 2013Class IIZ-0816-2014ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable…
RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.
Jul 5, 2013Class IIZ-1075-2014pCO2 D788 Membranes. Distributed by Radiometer America, Inc., Westlake, OH. An…
RADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.
Jun 10, 2013Class IIZ-0713-2014pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2…
RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This…
Nov 27, 2012Class IIZ-1930-2013TCM4 series Base Units, and spare part Power Supply, Radiometer Medical ApS, Akandevej 21…
RADIOMETER became aware that the TCM4 series Base Unit does not comply with IEC 60601-1 Sub Clause 15B.
Oct 8, 2012Class IIZ-2303-2013Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in…
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative…
Jul 5, 2012Class IIZ-2162-2012The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the… and 2 more products
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5…

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

11 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1046-2015, this FEI as the recalling firmDec 8, 201410 years or older, so it adds no points0
Class II recall Z-1293-2015, this FEI as the recalling firmOct 29, 201410 years or older, so it adds no points0
Class II recall Z-0873-2015, this FEI as the recalling firmOct 20, 201410 years or older, so it adds no points0
Class II recall Z-2597-2014, this FEI as the recalling firmJul 28, 201410 years or older, so it adds no points0
Class II recall Z-0816-2014, this FEI as the recalling firmOct 15, 201310 years or older, so it adds no points0
Class II recall Z-1075-2014, this FEI as the recalling firmJul 5, 201310 years or older, so it adds no points0
Class II recall Z-0713-2014, this FEI as the recalling firmJun 10, 201310 years or older, so it adds no points0
Class II recall Z-1930-2013, this FEI as the recalling firmNov 27, 201210 years or older, so it adds no points0
Class II recall Z-2303-2013, this FEI as the recalling firmOct 8, 201210 years or older, so it adds no points0
Class II recall Z-2162-2012, this FEI as the recalling firmJul 5, 201210 years or older, so it adds no points0
Inspection classified VAIAug 18, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing RADIOMETER AMERICA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check RADIOMETER AMERICA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Radiometer America Inc
Location
Middleburg Heights, OH (US)
Source
device