RADIOMETER AMERICA, INC. FDA Inspection Record
RADIOMETER AMERICA, INC. is an FDA-registered facility (FEI 1523456) located in Middleburg Heights, OH, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 2, 2015: No Action Indicated (NAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 11, latest Dec 8, 2014
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 2, 2015 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Justification for not reporting (21 CFR 806.20(b)(4)) |
| Aug 18, 2011 | Voluntary Action Indicated (VAI) | Rationale documented for no investigation (21 CFR 820.198(b)) Servicing - Lack of or inadequate procedures (21 CFR 820.200(a)) Supplier oversight (21 CFR 820.50(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 8, 2014 | Class II | Z-1046-2015 | ABL90 FLEX analyzer; Model number 393-090. A portable, automated analyzer that measures… The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a… |
| Oct 29, 2014 | Class II | Z-1293-2015 | Crea A and B membranes, packaged 4 per box
Used in conjunction with the ABL800 FLEX… Negative drift on QC during in-use lifetime of CREA membranes can be observed. |
| Oct 20, 2014 | Class II | Z-0873-2015 | HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for… Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf… |
| Jul 28, 2014 | Class II | Z-2597-2014 | RADIANCE v. 2.4X, 2.5X and 2.6X Product Usage: RADIANCE is a data processing… |
| Oct 15, 2013 | Class II | Z-0816-2014 | ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21
Br¿nsh¿j, Denmark
A portable… RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results. |
| Jul 5, 2013 | Class II | Z-1075-2014 | pCO2 D788 Membranes. Distributed by Radiometer America, Inc., Westlake, OH.
An… RADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results. |
| Jun 10, 2013 | Class II | Z-0713-2014 | pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145
pO2… RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This… |
| Nov 27, 2012 | Class II | Z-1930-2013 | TCM4 series Base Units, and spare part Power Supply, Radiometer Medical ApS, Akandevej 21… RADIOMETER became aware that the TCM4 series Base Unit does not comply with IEC 60601-1 Sub Clause 15B. |
| Oct 8, 2012 | Class II | Z-2303-2013 | Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in… RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative… |
| Jul 5, 2012 | Class II | Z-2162-2012 | The ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the… and 2 more products A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
11 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1046-2015, this FEI as the recalling firm | Dec 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1293-2015, this FEI as the recalling firm | Oct 29, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0873-2015, this FEI as the recalling firm | Oct 20, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2597-2014, this FEI as the recalling firm | Jul 28, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0816-2014, this FEI as the recalling firm | Oct 15, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1075-2014, this FEI as the recalling firm | Jul 5, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0713-2014, this FEI as the recalling firm | Jun 10, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1930-2013, this FEI as the recalling firm | Nov 27, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2303-2013, this FEI as the recalling firm | Oct 8, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2162-2012, this FEI as the recalling firm | Jul 5, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 18, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing RADIOMETER AMERICA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check RADIOMETER AMERICA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Radiometer America Inc
- Location
- Middleburg Heights, OH (US)
- Source