APPLIED MEDICAL TECHNOLOGY, INC. FDA Inspection Record
APPLIED MEDICAL TECHNOLOGY, INC. is an FDA-registered facility (FEI 1526012) located in Brecksville, OH, US. RegDataLab tracks 2 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 3, 2012: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jun 8, 2011. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Oct 6, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 3, 2012 | Voluntary Action Indicated (VAI) | Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2)) Production processes (21 CFR 820.70(a)) Lack of System for Determining MDR Events (21 CFR 803.17(a)(2)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Apr 29, 2011 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design plans - Lack of or inadequate (21 CFR 820.30(b)) Design input - documentation (21 CFR 820.30(c)) Design review - documentation (21 CFR 820.30(e)) and 5 more |
Warning letters
- Jun 8, 2011: Warning letter listed on FDA's Data Dashboard. View compliance record (193052).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 6, 2025 | Class II | Z-0494-2026 | Applied Medical Technology, Inc., NutraGlide, 5 f - 6F x 55 Nasal Feeding Tube with Stylet… and 4 more products The nasal feeding tube may be occluded |
| Aug 12, 2025 | Class II | Z-0127-2026 | NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 6 F x 55, REF E… and 3 more products The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces. |
| Nov 26, 2024 | Class II | Z-0819-2025 | AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Model/Catalog… The extension sets subject to this recall have ENFit connections, although they should have non- ENFit connections. ENFit connectors will… |
| May 14, 2024 | Class II | Z-2238-2024 | AMT Initial Placement Dilator Set. Used to place gastrostomy devices. Devices were labelled with the incorrect guidewire labelling. |
| Nov 10, 2020 | Class III | Z-0698-2021 | AMT Male ENFit-to-Stepped Male (Christmas Tree) Adapter, Model Number ETRN102.
Box PN… The products subject to this recall may contain obvious, incorrect pouch labels while the outer box has the correct label. |
| Jan 19, 2017 | Class III | Z-1223-2017 | AMT Bridle - Nasal Tube Retaining System
Used with Nasogastric/Nasointestinal (NG/NI)… and 1 more product Expired expiration dates are listed on the pouches. |
| Feb 17, 2016 | Class II | Z-1608-2016 | TLC¿ Self Retaining Retractor System Contents 1-RETRACTOR FRAME 6-ELASTIC STAYS STERILE EO… and 3 more products The firm received a complaint of an open pouch which breaches the sterilization validation. |
| Jan 26, 2016 | Class II | Z-1497-2016 | AMT G-Jet¿ Low-Profile Transgastric-Jejunal Feeding Device 14F 1.5cm 45cm SINGLE USE ONLY… The firm received complaints that the product is shorter in length than indicated on the label. |
FDA record score
14 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0494-2026, this FEI as the recalling firm | Oct 6, 2025 | 4 x 1 | 4 |
| Class II recall Z-0127-2026, this FEI as the recalling firm | Aug 12, 2025 | 4 x 1 | 4 |
| Class II recall Z-0819-2025, this FEI as the recalling firm | Nov 26, 2024 | 4 x 1 | 4 |
| Class II recall Z-2238-2024, this FEI as the recalling firm | May 14, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1608-2016, this FEI as the recalling firm | Feb 17, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1497-2016, this FEI as the recalling firm | Jan 26, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 3, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jun 8, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing APPLIED MEDICAL TECHNOLOGY, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check APPLIED MEDICAL TECHNOLOGY, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Applied Medical Technology Inc
- Location
- Brecksville, OH (US)
- Source
- device