THE DAAVLIN DISTRIBUTING CO. FDA Inspection Record
THE DAAVLIN DISTRIBUTING CO. is an FDA-registered facility (FEI 1526255) located in Bryan, OH, US. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Feb 3, 2012: No Action Indicated (NAI)
- Inspections on record
- 1 (1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Mar 5, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 3, 2012 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 5, 2026 | Class II | Z-2742-2026 | Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders and 2 more products Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7"… |
| Dec 10, 2025 | Class II | Z-2295-2026 | Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders… Some of the Calibration/Output Certificates had the outputs reversed. |
| Apr 26, 2022 | Class II | Z-1073-2022 | 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software… and 1 more product Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting… |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2742-2026, this FEI as the recalling firm | Mar 5, 2026 | 4 x 1 | 4 |
| Class II recall Z-2295-2026, this FEI as the recalling firm | Dec 10, 2025 | 4 x 1 | 4 |
| Class II recall Z-1073-2022, this FEI as the recalling firm | Apr 26, 2022 | 4 x 0.6 | 2.4 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing THE DAAVLIN DISTRIBUTING CO. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.