DEPUY SPINE, INC. FDA Inspection Record

FEI 1526439 · Inspections, Form 483s, warning letters and import refusals

DEPUY SPINE, INC. is an FDA-registered facility (FEI 1526439) located in RAYNHAM, MA, US. RegDataLab tracks 7 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Apr 18, 2019: No Action Indicated (NAI)
Inspections on record
7 (3 VAI, 4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
11, latest May 20, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 18, 2019No Action Indicated (NAI)No citations listed
Jun 16, 2017No Action Indicated (NAI)No citations listed
Jun 11, 2012No Action Indicated (NAI)No citations listed
Feb 10, 2012Voluntary Action Indicated (VAI)Sponsors' general responsibilities (21 CFR 812.40)
Sponsor securing investigator compliance (21 CFR 812.46(a))
Apr 28, 2009Voluntary Action Indicated (VAI)Production processes (21 CFR 820.70(a))
Apr 21, 2009No Action Indicated (NAI)No citations listed
Mar 11, 2009Voluntary Action Indicated (VAI)Sponsors' general responsibilities (21 CFR 812.40)
Sponsor securing investigator compliance (21 CFR 812.46(a))
Sponsor progress reports for significant risk study (21 CFR 812.150(b)(5))

Recalls

Showing the 10 most recent of 11 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 20, 2022Class IIZ-1323-2022EIT CIF cage, H 6mm, 8degree, L- an intervertebral body fusion devices indicated for use… and 1 more product
Label on the Outer Carton does not match the labels on the Outer Blister package, Inner Blister package and Patient Labels. The only label…
Apr 8, 2019Class IIZ-1225-2019Concorde Lift, Expandable Interbody Device, spinal fusion implant Convex 9x21… and 1 more product
A complaint trend was observed for post-operative loss of height and migration for Concorde Lift G3+ implants.
Dec 15, 2017Class IIZ-1300-2018Confidence Plus Kit Spinal Cement System, Product Codes: 283910000 (GTIN…
Certain lots of the spinal cement may contain incorrect amounts of contrast medium so that the product is difficult or impossible to…
Oct 12, 2016Class IIZ-0624-2017DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x… and 6 more products
SKYLINE Variable Ti screws may have contained SKYLINE Constrained Ti screws within the packaging.
Nov 20, 2015Class IIZ-0454-2016Aegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the…
Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The…
Apr 21, 2015Class IIZ-1674-2015EXPEDIUM Spine System/Synapse System; MODEL: EXPEDIUM Offset Conical Lock Nut 8mm, CATALOG…
Device does not mate as intended.
Jul 31, 2013Class IIZ-0093-2014PIPELINE LS ACCESS SYSTEM, 15 Dilator, 5 mm, Insulated The PIPLINE LS Dilators and… and 4 more products
Dilators and Holding Clip are being used for a secondary function of neuromonitoring, for which it is not 510(k) cleared.
Apr 17, 2013Class IIZ-2132-2013CONFIDENCE SPINAL CEMENT SYSTEM 5CC KIT, Product Code: 2839-05-000 The CONFIDENCE… and 2 more products
During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and…
Jan 15, 2013Class IIZ-0960-2013DePuy CONFIDENCE SPINAL CEMENT SYSTEM 5CC KIT Catalog Number:283905000. Delivery of… and 3 more products
Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck
Nov 7, 2012Class IIZ-0705-2013DePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal… and 8 more products
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque…

FDA record score

4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

14 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1323-2022, this FEI as the recalling firmMay 20, 20224 x 0.62.4
Class II recall Z-1225-2019, this FEI as the recalling firmApr 8, 20194 x 0.251
Class II recall Z-1300-2018, this FEI as the recalling firmDec 15, 20174 x 0.251
Class II recall Z-0624-2017, this FEI as the recalling firmOct 12, 201610 years or older, so it adds no points0
Class II recall Z-0454-2016, this FEI as the recalling firmNov 20, 201510 years or older, so it adds no points0
Class II recall Z-1674-2015, this FEI as the recalling firmApr 21, 201510 years or older, so it adds no points0
Class II recall Z-0093-2014, this FEI as the recalling firmJul 31, 201310 years or older, so it adds no points0
Class II recall Z-2132-2013, this FEI as the recalling firmApr 17, 201310 years or older, so it adds no points0
Class II recall Z-0960-2013, this FEI as the recalling firmJan 15, 201310 years or older, so it adds no points0
Class II recall Z-0705-2013, this FEI as the recalling firmNov 7, 201210 years or older, so it adds no points0
Class II recall Z-0685-2013, this FEI as the recalling firmJul 24, 201210 years or older, so it adds no points0
Inspection classified VAIFeb 10, 201210 years or older, so it adds no points0
Inspection classified VAIApr 28, 200910 years or older, so it adds no points0
Inspection classified VAIMar 11, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing DEPUY SPINE, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check DEPUY SPINE, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
DePuy Spine, Inc.
Location
RAYNHAM, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.