Western Enterprises, Inc. FDA Inspection Record
Western Enterprises, Inc. is an FDA-registered facility (FEI 1526809) located in Westlake, OH, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Aug 29, 2018: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (3 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Oct 10, 2014. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Jan 3, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 29, 2018 | Voluntary Action Indicated (VAI) | Documentation - review in response to changes or deviations (21 CFR 820.75(c)) Documentation (21 CFR 820.100(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Aug 12, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Investigation of device failures (21 CFR 820.198(c)) Documentation (21 CFR 820.100(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 1 more |
| Sep 2, 2014 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 3 more |
| May 27, 2010 | Voluntary Action Indicated (VAI) | Training records (21 CFR 820.25(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
Warning letters
- Oct 10, 2014: Warning letter listed on FDA's Data Dashboard. View compliance record (441366).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 3, 2023 | Class II | Z-1376-2023 | Digital VIPR
ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only)
ROC-9840… Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to… |
| Oct 16, 2021 | Class II | Z-0267-2022 | Roberts Digital VIPR System (with E-size cylinder)
Model: ROC-9840 and 3 more products VIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and… |
| Aug 15, 2019 | Class II | Z-0339-2020 | Grab n Go Opti series VIPR system
Model # PRX-9615 - Product Usage: The Grab n Go Opti… and 4 more products Limited access to flow settings as a result of the control knob having been rotated beyond its functional range |
| Jan 31, 2015 | Class I | Z-2069-2015 | Various OxyTOTE Portable Oxygen System regulators, Valve Integrated Pressure Regulator… The firm has become aware of the possibility of ignition and rupture (explosion) of a compressed gas oxygen cylinder, which can cause… |
| Apr 7, 2014 | Class II | Z-1849-2014 | Model RG3000 and RG4000 Series Valve Integrated Pressure Regulator (VIPR) Separation events have occurred, including the VIPR separating from the Aluminum cylinder. |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1376-2023, this FEI as the recalling firm | Jan 3, 2023 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Aug 29, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-0267-2022, this FEI as the recalling firm | Oct 16, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0339-2020, this FEI as the recalling firm | Aug 15, 2019 | 4 x 0.25 | 1 |
| Inspection classified VAI | Aug 12, 2016 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-2069-2015, this FEI as the recalling firm | Jan 31, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Oct 10, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1849-2014, this FEI as the recalling firm | Apr 7, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 27, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Western Enterprises, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Western Enterprises, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Western/Scott Fetzer Company
- Location
- Westlake, OH (US)
- Source
- device