FUJIFILM Healthcare Americas Corporation FDA Inspection Record
FUJIFILM Healthcare Americas Corporation is an FDA-registered facility (FEI 1528028) located in Twinsburg, OH, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 6, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (3 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 19, latest Oct 14, 2021
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 6, 2017 | Voluntary Action Indicated (VAI) | Complaints (21 CFR 820.198(a)) Documentation (21 CFR 820.100(b)) Report of risk to health (21 CFR 806.10(a)(1)) |
| Apr 8, 2013 | Voluntary Action Indicated (VAI) | Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1)) Documentation (21 CFR 820.100(b)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Feb 1, 2010 | Voluntary Action Indicated (VAI) | Quality Audit/Reaudit - conducted (21 CFR 820.22) Documented evaluation (21 CFR 820.50(a)(1)) Report of risk to health (21 CFR 806.10(a)(1)) Time to report - 10 days (21 CFR 806.10(b)) and 1 more |
| Aug 4, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 14, 2021 | Class II | Z-0189-2022 | Arietta 750 Ultrasound
Software Version: V1.0.0 through V2.1.3 and 2 more products Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does… |
| Oct 14, 2021 | Class II | Z-0192-2022 | Arietta 65 Ultrasound
Software Versions: V1.0 through V4.0.0 When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation… |
| Jul 29, 2021 | Class II | Z-2337-2021 | ECHELON MRI System and 2 more products There was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is… |
| Jun 30, 2021 | Class II | Z-2308-2021 | Model L43K Intraoperative Probe-For the diagnostic ultrasound evaluation… The screw cover at the tip of the probe may fall off after a cleaning and sterilization process. Users advised to conduct a preventive… |
| Mar 1, 2021 | Class II | Z-1282-2021 | Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained… Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function… |
| Jan 31, 2020 | Class II | Z-1341-2020 | UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is… These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings. |
| Oct 16, 2019 | Class II | Z-0271-2020 | Hitachi Scenaria Whole-body X-ray CT System There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach… |
| Jun 29, 2018 | Class II | Z-2839-2018 | Hitachi Scenaria CT system
Product Usage:
The SCENARIA CT system is indicated for… The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table… |
| Jun 18, 2018 | Class II | Z-2479-2018 | Hitachi Oasis MRI system
Product Usage:
Hitachi MR system is an imaging device and is… The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet… |
| Oct 20, 2017 | Class II | Z-0816-2018 | Ultrasound Transducer, Model Number: UST-5550-R, is used in conjunction with a diagnostic… The ultrasound probe may not have adequate protection against electrical shock hazards. |
FDA record score
14 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
22 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Feb 6, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0189-2022, this FEI as the recalling firm | Oct 14, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0192-2022, this FEI as the recalling firm | Oct 14, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2337-2021, this FEI as the recalling firm | Jul 29, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2308-2021, this FEI as the recalling firm | Jun 30, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1282-2021, this FEI as the recalling firm | Mar 1, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1341-2020, this FEI as the recalling firm | Jan 31, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0271-2020, this FEI as the recalling firm | Oct 16, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2839-2018, this FEI as the recalling firm | Jun 29, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2479-2018, this FEI as the recalling firm | Jun 18, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0816-2018, this FEI as the recalling firm | Oct 20, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2090-2017, this FEI as the recalling firm | May 2, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1096-2017, this FEI as the recalling firm | Jan 6, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2392-2016, this FEI as the recalling firm | Jun 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1540-2017, this FEI as the recalling firm | Apr 16, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2564-2014, this FEI as the recalling firm | Aug 18, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2366-2015, this FEI as the recalling firm | Apr 25, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 8, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1559-2013, this FEI as the recalling firm | Jan 31, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2254-2012, this FEI as the recalling firm | Jun 15, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2037-2012, this FEI as the recalling firm | May 4, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 1, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing FUJIFILM Healthcare Americas Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check FUJIFILM Healthcare Americas Corporation yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- FujiFilm Healthcare Americas Corporation
- Location
- Twinsburg, OH (US)
- Source
- device