HELENA LABORATORIES, CORP. FDA Inspection Record

FEI 1618982 · Inspections, Form 483s, warning letters and import refusals

HELENA LABORATORIES, CORP. is an FDA-registered facility (FEI 1618982) located in BEAUMONT, TX, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 14, 2018: Voluntary Action Indicated (VAI)
Inspections on record
4 (1 OAI, 3 VAI)
Last OAI inspection
Jul 21, 2017
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
15, latest Sep 14, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 14, 2018Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Lack of or inadequate process validation (21 CFR 820.75(a))
Jul 21, 2017Official Action Indicated (OAI)Documentation (21 CFR 820.75(a))
Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a))
and 4 more
Mar 1, 2013Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Design plans - Lack of or inadequate (21 CFR 820.30(b))
Design input - documentation (21 CFR 820.30(c))
and 5 more
Oct 29, 2010Voluntary Action Indicated (VAI)No citations listed

Recalls

Showing the 10 most recent of 15 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Sep 14, 2026Class IIZ-0469-2027V8 Serum Protein SPE Kit, Cat. No.: 1805, containing V8 SPE Diluent, Product No.: 552512
Serum protein kit has a diluent that may produce a decrease in the albumin fraction with a corresponding increase in the beta and gamma…
May 2, 2023Class IIZ-1975-2023Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus…
Packaging for a serrated blade applicator kit (12 Sample) may contain a serrated blade applicator (18 sample). This results in a…
Apr 6, 2022Class IIZ-1244-2022V8 Immunodisplacement Kit REF 1803
Due to microbial growth causing interference with interpretation.
Jan 5, 2021Class IIZ-1108-2021Cascade Abrazo aPTT Test Card - Product Usage: are to be used with the Cascade Abrazo…
Intermittent potential for shorter than expected clot times.
Nov 9, 2020Class IIZ-0635-2021K-ACT Actalyke Clotting Test Tubes that are used in the: ACTALYKE XL ACTIVATED CLOTTING…
Due to complaints received associated with cracked/split test tube caps.
Aug 8, 2018Class IIZ-1949-2019ColoCARE, Cat. No. 5650, packaged 250 envelopes/box, For In-Vitro Diagnostic Use. The… and 1 more product
The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning…
May 23, 2018Class IIIZ-0536-2020Platelet Aggregation Kit, Cat. No.5369, For use in platelet aggregation studies, Contains… and 1 more product
Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet Aggregation Kit…
Feb 27, 2018Class IIIZ-0297-2020SPIFE Alkaline Hemoglobin Kit, Cat. No. 3415, for use in the separation and…
Lots of gel kits contained a newer revision of the Customer Procedures instructions than intended. The incorrect inserts refer to updated…
Dec 7, 2017Class IIZ-2761-2019Cascade Abrazo aPTT Test Card, Model 5722, For Export Only. The firm name on the label is…
Some of the cards show occasional shorter clot times than the expected variation resulting in a potential false negative result on the…
Jul 21, 2017Class IIZ-0045-2020SPIFE ImmunoFix-6 Kit, Cat. No. 3401, 3401T, For In-Vitro Diagnostic Use. The firm name…
The plates do not have wells punched for the controls in the patient 1 area.

FDA record score

25 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

16 records behind this number
RecordDatePoints x age weightAdds
Inspection classified OAIJul 21, 201730 x 0.257.5
Class II recall Z-0469-2027, this FEI as the recalling firmSep 14, 20264 x 14
Class II recall Z-1975-2023, this FEI as the recalling firmMay 2, 20234 x 0.62.4
Class II recall Z-1244-2022, this FEI as the recalling firmApr 6, 20224 x 0.62.4
Inspection classified VAIDec 14, 20186 x 0.251.5
Class II recall Z-1108-2021, this FEI as the recalling firmJan 5, 20214 x 0.251
Class II recall Z-0635-2021, this FEI as the recalling firmNov 9, 20204 x 0.251
Class II recall Z-1949-2019, this FEI as the recalling firmAug 8, 20184 x 0.251
Class II recall Z-2761-2019, this FEI as the recalling firmDec 7, 20174 x 0.251
Class II recall Z-0045-2020, this FEI as the recalling firmJul 21, 20174 x 0.251
Class II recall Z-0106-2019, this FEI as the recalling firmApr 4, 20174 x 0.251
Class II recall Z-0561-2019, this FEI as the recalling firmFeb 24, 20174 x 0.251
Class II recall Z-2434-2019, this FEI as the recalling firmOct 14, 201610 years or older, so it adds no points0
Class II recall Z-0594-2019, this FEI as the recalling firmApr 20, 201610 years or older, so it adds no points0
Inspection classified VAIMar 1, 201310 years or older, so it adds no points0
Inspection classified VAIOct 29, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing HELENA LABORATORIES, CORP. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check HELENA LABORATORIES, CORP. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Helena Laboratories, Corp.
Location
BEAUMONT, TX (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.