B Braun Medical, Inc. FDA Inspection Record

FEI 1641965 · Inspections, Form 483s, warning letters and import refusals

B Braun Medical, Inc. is an FDA-inspected facility (FEI 1641965) located in Carrollton, TX, US. RegDataLab tracks 2 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Aug 16, 2012: Official Action Indicated (OAI)
Inspections on record
2 (2 OAI)
Last OAI inspection
Aug 16, 2012
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Nov 23, 2020

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 16, 2012Official Action Indicated (OAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Availability (21 CFR 820.180)
Investigation of device failures (21 CFR 820.198(c))
Report of risk to health (21 CFR 806.10(a)(1))
and 2 more
Oct 8, 2010Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Incorrect translation to production specifications (21 CFR 820.30(h))
Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1))
Documentation (21 CFR 820.100(b))
and 1 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Nov 23, 2020Class IIZ-0717-2021APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product… and 2 more products
There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble…
Dec 5, 2012Class IIZ-2175-2013Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used…
B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the…
Nov 14, 2012Class IZ-2188-2013Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used…
This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in…
Oct 31, 2012Class IIZ-2164-2013Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used…
This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in…
Oct 24, 2012Class IIZ-2163-2013Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used…
B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as…

FDA record score

1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0717-2021, this FEI as the recalling firmNov 23, 20204 x 0.251
Class II recall Z-2175-2013, this FEI as the recalling firmDec 5, 201210 years or older, so it adds no points0
Class I recall Z-2188-2013, this FEI as the recalling firmNov 14, 201210 years or older, so it adds no points0
Class II recall Z-2164-2013, this FEI as the recalling firmOct 31, 201210 years or older, so it adds no points0
Class II recall Z-2163-2013, this FEI as the recalling firmOct 24, 201210 years or older, so it adds no points0
Inspection classified OAIAug 16, 201210 years or older, so it adds no points0
Inspection classified OAIOct 8, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing B Braun Medical, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check B Braun Medical, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
B Braun Medical, Inc.
Location
Carrollton, TX (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.