B Braun Medical, Inc. FDA Inspection Record
B Braun Medical, Inc. is an FDA-inspected facility (FEI 1641965) located in Carrollton, TX, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 16, 2012: Official Action Indicated (OAI)
- Inspections on record
- 2 (2 OAI)
- Last OAI inspection
- Aug 16, 2012
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Nov 23, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 16, 2012 | Official Action Indicated (OAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Availability (21 CFR 820.180) Investigation of device failures (21 CFR 820.198(c)) Report of risk to health (21 CFR 806.10(a)(1)) and 2 more |
| Oct 8, 2010 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Incorrect translation to production specifications (21 CFR 820.30(h)) Evaluation and Selection, Suppliers, Contractors, etc. (21 CFR 820.50(a)(1)) Documentation (21 CFR 820.100(b)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 23, 2020 | Class II | Z-0717-2021 | APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product… and 2 more products There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble… |
| Dec 5, 2012 | Class II | Z-2175-2013 | Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used… B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the… |
| Nov 14, 2012 | Class I | Z-2188-2013 | Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used… This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in… |
| Oct 31, 2012 | Class II | Z-2164-2013 | Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used… This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in… |
| Oct 24, 2012 | Class II | Z-2163-2013 | Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used… B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as… |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0717-2021, this FEI as the recalling firm | Nov 23, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2175-2013, this FEI as the recalling firm | Dec 5, 2012 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-2188-2013, this FEI as the recalling firm | Nov 14, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2164-2013, this FEI as the recalling firm | Oct 31, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2163-2013, this FEI as the recalling firm | Oct 24, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 16, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 8, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing B Braun Medical, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check B Braun Medical, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- B Braun Medical, Inc.
- Location
- Carrollton, TX (US)
- Source
- inspections