Edwards Lifesciences LLC FDA Inspection Record
Edwards Lifesciences LLC is an FDA-registered facility (FEI 1713910) located in Draper, UT, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Nov 19, 2021: No Action Indicated (NAI)
- Inspections on record
- 8 (1 OAI, 2 VAI, 5 NAI)
- Last OAI inspection
- Feb 22, 2013
- Warning letters
- 1 listed by FDA for this FEI, latest May 24, 2013
- Form 483 and inspection documents
- 1 in FDA's reading room
- Recalls as the recalling firm
- 21, latest Jan 22, 2019
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 19, 2021 | No Action Indicated (NAI) | No citations listed |
| Feb 1, 2018 | No Action Indicated (NAI) | No citations listed |
| May 20, 2016 | Voluntary Action Indicated (VAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) |
| Nov 20, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Nonconforming product control (21 CFR 820.90(a)) DHR content (21 CFR 820.184) Lack of or inadequate DHR procedures (21 CFR 820.184) and 1 more |
| Apr 11, 2014 | No Action Indicated (NAI) | No citations listed |
| Feb 22, 2013 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Nonconforming product control (21 CFR 820.90(a)) and 6 more |
| Jul 13, 2011 | No Action Indicated (NAI) | No citations listed |
| Nov 20, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 22, 2019 | Class II | Z-0835-2019 | Edward Lifesciences, Edwards Commander Delivery System:
Model Number UDI… This action has been initiated to address complaint reports received for crimp balloon and inflation balloon component separation during… |
| Mar 30, 2018 | Class II | Z-1929-2018 | Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile… Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon… |
| Jul 21, 2017 | Class I | Z-0293-2018 | Certitude Delivery System and any associated Multipacks Models - 9600CT20A, 9600CT23A… A component in the Certitude Loader may have a molding overflow defect that could detach during advancement of the delivery system… |
| Nov 4, 2015 | Class II | Z-0398-2016 | Thin-Flex Venous Cannula, Model Number:TF292902A Edwards Lifesciences initiated a field action for the Thin-Flex Venous Cannula (Model TF292902A) because the connector size is incorrectly… |
| Aug 26, 2015 | Class II | Z-2780-2015 | ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the… The configuration of the slot at the tip of the knot pusher may inhibit intended ease of use. Fraying or splitting of the suture could… |
| May 5, 2015 | Class II | Z-1758-2015 | Edwards Lifesciences IntraClude Intra-Aortic Occlusion Device, model number ICF100, 10.5… Potential device damage on the strain relief near the hub area, which may result in blood leakage. |
| Mar 23, 2015 | Class II | Z-1371-2015 | Edwards Lifesciences Fem-Flex II Femoral Arterial Cannula 8, 10, 12 French,Sterile EO, Rx… Edwards Lifesciences is recalling Fem-Flex II Pediatric Femoral Arterial Cannula sizes 8, 10, 12 French because of the potential of tissue… |
| Oct 8, 2014 | Class II | Z-0217-2015 | Edwards Lifesciences Vent Catheters, . Packaged sterile and non-pyrogenic in a sealed… Edwards Lifesciences is conducting a field action on the Vent Catheters due to the potential for disruption of the white plastic coating on… |
| Apr 9, 2014 | Class II | Z-1466-2014 | Edwards Lifesciences Fem-Flex II Arterial and femoral Cannulae, arterial cannula sizes… and 1 more product Potential for separation of dilator tips. |
| Jan 20, 2014 | Class II | Z-0918-2014 | Crimper Model 9100CR26
The Crimper is indicated for use in preparing the Edwards SAPIEN… Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when… |
Form 483s and inspection documents
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
24 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0293-2018, this FEI as the recalling firm | Jul 21, 2017 | 15 x 0.25 | 3.8 |
| Class II recall Z-0835-2019, this FEI as the recalling firm | Jan 22, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1929-2018, this FEI as the recalling firm | Mar 30, 2018 | 4 x 0.25 | 1 |
| Inspection classified VAI | May 20, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0398-2016, this FEI as the recalling firm | Nov 4, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2780-2015, this FEI as the recalling firm | Aug 26, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1758-2015, this FEI as the recalling firm | May 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1371-2015, this FEI as the recalling firm | Mar 23, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0217-2015, this FEI as the recalling firm | Oct 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1466-2014, this FEI as the recalling firm | Apr 9, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0918-2014, this FEI as the recalling firm | Jan 20, 2014 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0091-2014, this FEI as the recalling firm | Sep 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0134-2014, this FEI as the recalling firm | Aug 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1963-2013, this FEI as the recalling firm | Jul 30, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1972-2013, this FEI as the recalling firm | Jul 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1749-2013, this FEI as the recalling firm | Jun 19, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 24, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1542-2013, this FEI as the recalling firm | May 14, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1134-2013, this FEI as the recalling firm | Mar 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1111-2013, this FEI as the recalling firm | Mar 14, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 22, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1322-2013, this FEI as the recalling firm | Jan 25, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1967-2012, this FEI as the recalling firm | Jun 6, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Edwards Lifesciences LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Edwards Lifesciences LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Edwards Lifesciences, LLC
- Location
- Draper, UT (US)
- Source
- device