ELITechGroup Inc FDA Inspection Record
ELITechGroup Inc is an FDA-registered facility (FEI 1717966) located in LOGAN, UT, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 11, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (4 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 11, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Software validation for automated processes (21 CFR 820.70(i)) Equipment suitability & capability (21 CFR 820.72(a)) and 3 more |
| May 27, 2015 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) Documentation (21 CFR 820.100(b)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Jul 22, 2011 | Voluntary Action Indicated (VAI) | Design output - Lack of or inadequate procedures (21 CFR 820.30(d)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design review - documentation (21 CFR 820.30(e)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) and 3 more |
| Jul 1, 2009 | Voluntary Action Indicated (VAI) | Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Aug 11, 2017 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | May 27, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 22, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 1, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ELITechGroup Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ELITechGroup Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- ELITechGroup, Inc.
- Location
- LOGAN, UT (US)
- Source
- device