LivaNova USA, INC. FDA Inspection Record
LivaNova USA, INC. is an FDA-registered facility (FEI 1718850) located in ARVADA, CO, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Mar 30, 2016: No Action Indicated (NAI)
- Inspections on record
- 4 (1 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Dec 29, 2015. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 30, latest Apr 24, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 30, 2016 | No Action Indicated (NAI) | Sampling plans (21 CFR 820.250(b)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Documentation (21 CFR 820.100(b)) |
| Sep 1, 2015 | No Action Indicated (NAI) | No citations listed |
| Mar 26, 2013 | Voluntary Action Indicated (VAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design validation - production units (21 CFR 820.30(g)) |
| Jul 14, 2009 | No Action Indicated (NAI) | No citations listed |
Warning letters
- Dec 29, 2015: Warning letter listed on FDA's Data Dashboard. View compliance record (484629).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 24, 2020 | Class II | Z-2443-2020 | Sterile Disposable Connectors ASY CONN Y 1/2X1/2X3/8 STR 24. REF/ GTIN for Insert… and 21 more products The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The… |
| Feb 24, 2020 | Class II | Z-1498-2020 | LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to… The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM… |
| Nov 18, 2019 | Class III | Z-2642-2020 | LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk… |
| Jul 3, 2019 | Class III | Z-1389-2020 | LivaNova SMART PERFUSION PACK INSPIRE 8S - Product Usage: The affected product is a… Perfusion tubing pack was packaged with the incorrect cuvette. |
| Jan 28, 2019 | Class III | Z-1119-2020 | LivaNova Sorin Connect DPIII KOMPL. S5 REF 96-415-130, 2017-08-15, IPX1, RXOnly… Incorrect labeling; it was identified that some Datapad devices contained incorrect labels. Specifically the manufacturing date in the… |
| Nov 9, 2018 | Class II | Z-0770-2019 | Perfusion Pack, Rx only, Sterile EO 4-way stopcocks and stopcock manifold assembled in certain perfusion tubing systems, indicated for use in surgical procedures requiring… |
| Nov 1, 2018 | Class II | Z-0220-2020 | SORIN Heater-Cooler System 3T 120V~/60Hz
REF 16-02-85
Description of product:
The… The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit… |
| Sep 14, 2018 | Class II | Z-1099-2019 | SORIN Heater-Cooler System 3T, REF 16-02-85
Intended for use with a Stockert/Sorin… For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM). |
| Apr 20, 2018 | Class II | Z-1901-2018 | Sorin Stockert Heater-Cooler System 3T,
a) Item #: 16-02-81, 240 v/60 Hz;
b) Item… Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices… |
| Mar 13, 2018 | Class II | Z-1918-2018 | Inspire 6 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050713
The… and 1 more product Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar… |
FDA record score
11 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
27 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2443-2020, this FEI as the recalling firm | Apr 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1498-2020, this FEI as the recalling firm | Feb 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0770-2019, this FEI as the recalling firm | Nov 9, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0220-2020, this FEI as the recalling firm | Nov 1, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1099-2019, this FEI as the recalling firm | Sep 14, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1901-2018, this FEI as the recalling firm | Apr 20, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1918-2018, this FEI as the recalling firm | Mar 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0076-2018, this FEI as the recalling firm | Sep 28, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2771-2017, this FEI as the recalling firm | Jun 30, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2783-2017, this FEI as the recalling firm | Jun 29, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2871-2017, this FEI as the recalling firm | Jun 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0716-2017, this FEI as the recalling firm | Oct 31, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0412-2017, this FEI as the recalling firm | Sep 29, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2866-2016, this FEI as the recalling firm | Aug 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2756-2016, this FEI as the recalling firm | Aug 12, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2147-2016, this FEI as the recalling firm | Jun 22, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0813-2016, this FEI as the recalling firm | Jan 25, 2016 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Dec 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2076-2015, this FEI as the recalling firm | Jun 15, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0682-2017, this FEI as the recalling firm | Aug 1, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1985-2013, this FEI as the recalling firm | Jul 8, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0928-2013, this FEI as the recalling firm | Jan 28, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0166-2013, this FEI as the recalling firm | Sep 20, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0271-2013, this FEI as the recalling firm | Sep 19, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0157-2013, this FEI as the recalling firm | Sep 14, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1186-2016, this FEI as the recalling firm | Aug 15, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing LivaNova USA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check LivaNova USA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- LivaNova USA Inc.
- Location
- ARVADA, CO (US)
- Source