LUMENIS BE INC. FDA Inspection Record
LUMENIS BE INC. is an FDA-registered facility (FEI 1720381) located in Salt Lake City, UT, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 30, 2018: No Action Indicated (NAI)
- Inspections on record
- 4 (1 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Oct 14, 2021
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 30, 2018 | No Action Indicated (NAI) | No citations listed |
| Sep 13, 2016 | No Action Indicated (NAI) | Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Failure to report (21 CFR 1002.20(a)) |
| Mar 15, 2013 | No Action Indicated (NAI) | No citations listed |
| Aug 21, 2009 | Voluntary Action Indicated (VAI) | Software validation for automated processes (21 CFR 820.70(i)) Rationale documented for no investigation (21 CFR 820.198(b)) Analysis of data/reports from data sources (21 CFR 820.100(a)(1)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 14, 2021 | Class I | Z-0194-2022 | IPL Coupling Gel, I L (AX1009013), 12 pack of 0.25 L (AX1009018), 6 pack of 1 L (AX1009019… Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA… |
| Aug 15, 2014 | Class II | Z-2733-2014 | Array LaserLink, Manufactured by Lumenis,
The Array" LaserLink" is a laser system… Lumenis initiated a field-correction for the Array Laser Link", GA-0006700 (SN XXYYZZ) because the system may project an erroneous… |
| Aug 19, 2013 | Class II | Z-2191-2013 | Vision One Laser System Model GA-0025020, Serial No: 10050, 10051, 10053.
Intended for… Lumenis has initiated a recall on certain models of Vision One System due to a potential for unintended laser exposure to the user. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0194-2022, this FEI as the recalling firm | Oct 14, 2021 | 15 x 0.25 | 3.8 |
| Class II recall Z-2733-2014, this FEI as the recalling firm | Aug 15, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2191-2013, this FEI as the recalling firm | Aug 19, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 21, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing LUMENIS BE INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.