STRYKER CORP. FDA Inspection Record
STRYKER CORP. is an FDA-registered facility (FEI 1811755) located in KALAMAZOO, MI, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jan 7, 2016: No Action Indicated (NAI)
- Inspections on record
- 6 (2 OAI, 1 VAI, 3 NAI)
- Last OAI inspection
- Nov 9, 2012
- Warning letters
- 1 listed by FDA for this FEI, latest Mar 6, 2013
- Form 483 and inspection documents
- 2 in FDA's reading room
- Recalls as the recalling firm
- 55, latest Jun 15, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 7, 2016 | No Action Indicated (NAI) | No citations listed |
| Sep 13, 2013 | No Action Indicated (NAI) | No citations listed |
| Nov 9, 2012 | Official Action Indicated (OAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Equipment control activity documentation (21 CFR 820.72(a)) Supplier oversight (21 CFR 820.50(a)(2)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) and 2 more |
| Jun 18, 2012 | No Action Indicated (NAI) | Documentation (21 CFR 820.80(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) Documentation (21 CFR 820.100(b)) |
| Sep 24, 2010 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3)) Investigation of device failures (21 CFR 820.198(c)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) and 1 more |
| May 14, 2009 | Official Action Indicated (OAI) | Training (21 CFR 820.25(b)) Design validation - simulated testing (21 CFR 820.30(g)) Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b)) Specific non-conforming product procedures (21 CFR 820.90(a)) and 6 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 15, 2022 | Class II | Z-1452-2022 | SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090… When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a… |
| Oct 16, 2020 | Class II | Z-0431-2021 | Stryker Zyphr Disposable Cranial Perforator, Large 14/11 mm- is a sterile, single use… Potential for the inner bit to contain a crack, that may or may not be visible, which may lead to metal fragments entering the surgical site… |
| Mar 26, 2020 | Class II | Z-1807-2020 | Neptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The… Due to a distribution error, product from the affected lot was shipped expired. |
| Jan 22, 2020 | Class II | Z-1402-2022 | Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for… Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood… |
| Jan 13, 2020 | Class II | Z-1097-2020 | SafeAir Telescopic Smoke
Evacuation Pencil, Product No. 0703 046 004. The product is a… and 2 more products Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user. |
| Nov 6, 2019 | Class II | Z-0592-2020 | Color Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder)… and 1 more product There is a potential for the cuff to leak air leading to a loss of vascular occlusion. |
| Jul 31, 2019 | Class II | Z-0023-2020 | C2 Nerve Monitor, 4 Channel, 5140-508-240, UDI 07613327177329
Product Usage: The C2… and 1 more product Stryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device… |
| Mar 19, 2019 | Class II | Z-1215-2019 | Performance Series Sagittal Blade, Catalog Number 6125-127-100
Product Usage:
The… During manufacture, packaging materials for a small portion of the lot may not have been properly positioned, resulting in a potential… |
| Dec 4, 2018 | Class III | Z-1394-2022 | Triton Sponge System (Software), GAUS-3, Software Versions 4.6.5, 4.8.7, 4.9.6 Gauss Surgical is updating the Instructions for Use (IFU) for our Triton Sponge System in order to clarify proper technique for handling… |
| Feb 23, 2018 | Class II | Z-1500-2018 | T4 Pullover Toga, (S/M); Catalog Number: 0400-710-000
Sterile personal protection… and 10 more products Separation of material layers may occur, causing a potential risk of exposure to contaminants. |
Form 483s and inspection documents
FDA record score
15 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
55 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1452-2022, this FEI as the recalling firm | Jun 15, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0431-2021, this FEI as the recalling firm | Oct 16, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1807-2020, this FEI as the recalling firm | Mar 26, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1402-2022, this FEI as the recalling firm | Jan 22, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1097-2020, this FEI as the recalling firm | Jan 13, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0592-2020, this FEI as the recalling firm | Nov 6, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0023-2020, this FEI as the recalling firm | Jul 31, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1215-2019, this FEI as the recalling firm | Mar 19, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1500-2018, this FEI as the recalling firm | Feb 23, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1243-2018, this FEI as the recalling firm | Nov 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1216-2018, this FEI as the recalling firm | Oct 16, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2071-2017, this FEI as the recalling firm | Mar 10, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1886-2017, this FEI as the recalling firm | Feb 28, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0868-2017, this FEI as the recalling firm | Nov 30, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-0112-2017, this FEI as the recalling firm | Aug 17, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2630-2016, this FEI as the recalling firm | Jul 1, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2238-2016, this FEI as the recalling firm | Jun 24, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2204-2016, this FEI as the recalling firm | Jun 16, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2536-2016, this FEI as the recalling firm | Jun 14, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1939-2016, this FEI as the recalling firm | Apr 28, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1378-2016, this FEI as the recalling firm | Mar 22, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1379-2016, this FEI as the recalling firm | Mar 21, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1037-2016, this FEI as the recalling firm | Feb 4, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0417-2016, this FEI as the recalling firm | Nov 23, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2524-2015, this FEI as the recalling firm | Jul 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2102-2015, this FEI as the recalling firm | Jun 2, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2099-2015, this FEI as the recalling firm | May 22, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2062-2015, this FEI as the recalling firm | May 8, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1810-2015, this FEI as the recalling firm | May 4, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1671-2015, this FEI as the recalling firm | Apr 24, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1562-2015, this FEI as the recalling firm | Mar 26, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1066-2015, this FEI as the recalling firm | Jan 7, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0172-2015, this FEI as the recalling firm | Oct 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2156-2014, this FEI as the recalling firm | Jun 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1749-2014, this FEI as the recalling firm | May 22, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1715-2014, this FEI as the recalling firm | May 13, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1493-2014, this FEI as the recalling firm | Mar 26, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1247-2014, this FEI as the recalling firm | Mar 18, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0902-2014, this FEI as the recalling firm | Dec 30, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0630-2014, this FEI as the recalling firm | Dec 6, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0127-2014, this FEI as the recalling firm | Nov 4, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0280-2014, this FEI as the recalling firm | Nov 1, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0092-2014, this FEI as the recalling firm | Oct 2, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1740-2013, this FEI as the recalling firm | Jul 1, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1878-2013, this FEI as the recalling firm | Jun 24, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0986-2015, this FEI as the recalling firm | Jun 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1489-2013, this FEI as the recalling firm | May 23, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Mar 6, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Nov 9, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2134-2012, this FEI as the recalling firm | Jul 17, 2012 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-2061-2012, this FEI as the recalling firm | Jun 8, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1809-2012, this FEI as the recalling firm | Jun 1, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0454-2013, this FEI as the recalling firm | May 24, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 24, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | May 14, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing STRYKER CORP. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check STRYKER CORP. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Stryker Corporation
- Location
- KALAMAZOO, MI (US)
- Source
- device