JARDON EYE PROSTHETICS, INC. FDA Inspection Record
JARDON EYE PROSTHETICS, INC. is an FDA-registered facility (FEI 1824736) located in SOUTHFIELD, MI, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 2, 2026: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (4 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Form 483 and inspection documents
- 1 in FDA's reading room
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 2, 2026 | Voluntary Action Indicated (VAI) | Lack of document control (ISO 13485:2016 Clause 4.2.4) Management responsibilities and authorities not documented (ISO 13485:2016 Clause 5.5.1) Management review procedures not documented (ISO 13485:2016 Clause 5.6.1) Requirements for infrastructure maintenance activities not documented (ISO 13485:2016 Clause 6.3) and 4 more |
| Aug 31, 2017 | Voluntary Action Indicated (VAI) | Procedures not adequately established or maintained (21 CFR 820.40) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) and 4 more |
| Dec 12, 2013 | No Action Indicated (NAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Documented evaluation (21 CFR 820.50(a)(1)) |
| Oct 7, 2011 | Voluntary Action Indicated (VAI) | No citations listed |
| Jun 16, 2010 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Supplier notification of changes (21 CFR 820.50(b)) |
| Apr 30, 2009 | No Action Indicated (NAI) | Documentation (21 CFR 820.80(e)) DMR device specifications (21 CFR 820.181(a)) DHR content (21 CFR 820.184) Approval, inadequate purchasing data (21 CFR 820.50(b)) and 5 more |
Form 483s and inspection documents
FDA record score
8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 2, 2026 | 6 x 1 | 6 |
| Inspection classified VAI | Aug 31, 2017 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Oct 7, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 16, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing JARDON EYE PROSTHETICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check JARDON EYE PROSTHETICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Jardon Eye Prosthetics, Inc.
- Location
- SOUTHFIELD, MI (US)
- Source
- device