Skytron, LLC FDA Inspection Record
Skytron, LLC is an FDA-registered facility (FEI 1825014) located in Grand Rapids, MI, US. RegDataLab tracks 1 FDA inspection and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 6, 2015: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest May 19, 2026. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 10, latest Aug 1, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 6, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) |
Warning letters
- May 19, 2026: Medical Device/Adulterated/Misbranded/Lacks PMA and/or 510(k). Read the letter on FDA.gov.
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 1, 2023 | Class II | Z-2469-2023 | GS70 Salus Surgical Table, Product Code 18-070-70 and 1 more product When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on… |
| Jun 2, 2022 | Class II | Z-1309-2022 | Product Name: GCX Channel Mounting Accessory
Model/Catalog Number: H8-300-40-A Under certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel… |
| May 11, 2021 | Class II | Z-1861-2021 | Skyton Bracket Flatscreen, Revision 2 & 3-The FS32 is a single monitor mounting bracket… Failure of the FS32 Monitor Brackets from the mounting block could potentially loosen over time and eventually detach from the baseplate… |
| Sep 11, 2020 | Class II | Z-0071-2021 | VGDF-SKY (R4-010-06) VGA TO FIBER CONVERTER-Single Format Faceplate, a component of the… VGDF (VGA) Faceplate failure results in failure of video signal for SkyVision Linx 300, a secondary display. Video would continue on the… |
| Mar 10, 2020 | Class II | Z-0109-2021 | SkyVision SDS System - Video Integration
Model: GC-SDS-BSCB-200 - Product Usage: is a… Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the… |
| Mar 5, 2020 | Class II | Z-1569-2020 | Skytron Ergon 3 Series Skyboom Monitor Bracket-Flatscreen Bracket, single monitor mount… Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to… |
| Apr 18, 2019 | Class II | Z-1779-2019 | Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam
Product… Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure |
| Apr 6, 2016 | Class II | Z-1844-2016 | EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide… The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to… |
| Mar 28, 2016 | Class II | Z-1871-2016 | Ergon III Series Skyboom Monitor Bracket. Part # 3FCM1 Flatscreen Bracket, single monitor… Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket… |
| Dec 2, 2015 | Class II | Z-0655-2016 | Ergon 2 Articulating Equipment Boom Monitor Brackets
mounting point for display… The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall… |
FDA record score
40 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
12 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Warning letter, no close-out letter listed | May 19, 2026 | 30 x 1 | 30 |
| Class II recall Z-2469-2023, this FEI as the recalling firm | Aug 1, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1309-2022, this FEI as the recalling firm | Jun 2, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1861-2021, this FEI as the recalling firm | May 11, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0071-2021, this FEI as the recalling firm | Sep 11, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0109-2021, this FEI as the recalling firm | Mar 10, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1569-2020, this FEI as the recalling firm | Mar 5, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1779-2019, this FEI as the recalling firm | Apr 18, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1844-2016, this FEI as the recalling firm | Apr 6, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1871-2016, this FEI as the recalling firm | Mar 28, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0655-2016, this FEI as the recalling firm | Dec 2, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 6, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |