HORIBA INSTRUMENTS INCORPORATED FDA Inspection Record
HORIBA INSTRUMENTS INCORPORATED is an FDA-registered facility (FEI 1827821) located in Canton, MI, US. RegDataLab tracks 7 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 31, 2017: No Action Indicated (NAI)
- Inspections on record
- 7 (2 OAI, 1 VAI, 4 NAI)
- Last OAI inspection
- Jun 30, 2015
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest Aug 10, 2018
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 31, 2017 | No Action Indicated (NAI) | No citations listed |
| Jun 30, 2015 | Official Action Indicated (OAI) | No citations listed |
| Jan 24, 2013 | No Action Indicated (NAI) | No citations listed |
| Mar 9, 2012 | No Action Indicated (NAI) | No citations listed |
| Jan 28, 2011 | No Action Indicated (NAI) | No citations listed |
| Jan 15, 2010 | Official Action Indicated (OAI) | Design input - documentation (21 CFR 820.30(c)) DMR device specifications (21 CFR 820.181(a)) Design validation - production units (21 CFR 820.30(g)) Design validation - simulated testing (21 CFR 820.30(g)) and 5 more |
| Apr 8, 2009 | Voluntary Action Indicated (VAI) | Quality audit corrective action, reaudits {see also 820.100} (21 CFR 820.22) Design input procedures - appropriateness (21 CFR 820.30(c)) Design history file (21 CFR 820.30(j)) Investigation procedures (21 CFR 820.100(a)(2)) and 5 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 10, 2018 | Class III | Z-0389-2019 | Pointe Scientific AutoHDL/LDL Cholesterol Calibrator, Catalog Number H7545-CAL - Product… Vials of the autoHDL/LDL Calibrator were shipped without a vial label. |
| May 19, 2015 | Class II | Z-2487-2015 | Pointe Scientific G6PD Controls
Kit configuration: 6 x 0.5 ml (2 vials each of Level 1… A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after… |
| Apr 3, 2015 | Class II | Z-2063-2016 | Liquid stable 2 part Homocysteine Reagent, for in vitro quantitative determination of… Deterioration in performance of the Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT resulting in under recovery. The manufacturer has… |
| Mar 13, 2015 | Class III | Z-2764-2015 | Pointe Scientific autoHDL Reagent H7545
H7545-R1 in a bulk cube container or HDPE… Crystals may be observed in the R1 reagent. Crystals may be a result of the reagent being frozen during transport. The crystals do not… |
| Dec 12, 2014 | Class III | Z-2773-2015 | Pointe Scientific Creatinine Reagent
Catalog number C7539-150
in vitro… Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1… |
| Oct 23, 2014 | Class III | Z-2316-2017 | MedTest DX Assayed Human Multi Sera Control Level 1 (HNC200) and Level 2… MedTest DX Assayed Human Multi Sera Control Level 1 (HNC200) and Level 2 (HEC200) control values are outside of specification. |
| May 15, 2014 | Class III | Z-2121-2017 | Pointe Scientific Chemistry Controls Level I and II
LEVl: Amber glass vial w/PP white… The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the… |
| Mar 13, 2012 | Class II | Z-2586-2017 | Pointe Scientific Hitachi Total Bilirubin Rl Reagent
Product Usage:
Two part Chemistry… Decrease reactivity of R 1 component. |
| Feb 15, 2012 | Class III | Z-2116-2017 | Pointe Scientific Hitachi AST Reagent
R1 : PE plastic bottle w/PP plastic cap.
R2: PE… A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the… |
| Oct 19, 2011 | Class III | Z-2650-2017 | Pointe Scientific Liquid Creatine Kinase Reagent Set
Device is a kit comprised of two… Linear performance information in product insert does not match that listed in the approved premarket notification |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified OAI | Jun 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2487-2015, this FEI as the recalling firm | May 19, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2063-2016, this FEI as the recalling firm | Apr 3, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2586-2017, this FEI as the recalling firm | Mar 13, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2165-2017, this FEI as the recalling firm | Jul 8, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jan 15, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 8, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 17, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing HORIBA INSTRUMENTS INCORPORATED as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check HORIBA INSTRUMENTS INCORPORATED yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Horiba Instruments Incorporated
- Location
- Canton, MI (US)
- Source
- device