Panoramic Corp. FDA Inspection Record
Panoramic Corp. is an FDA-inspected facility (FEI 1832462) located in Fort Wayne, IN, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Aug 7, 2019: No Action Indicated (NAI)
- Inspections on record
- 4 (4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Oct 1, 2015. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Jul 14, 2017
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 7, 2019 | No Action Indicated (NAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Sep 7, 2016 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) and 1 more |
| Apr 24, 2015 | No Action Indicated (NAI) | No citations listed |
| Aug 13, 2009 | No Action Indicated (NAI) | No citations listed |
Warning letters
- Oct 1, 2015: Warning letter listed on FDA's Data Dashboard. View compliance record (472841).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 14, 2017 | Class II | Z-0517-2018 | PC-1000, PC-1000/Laser1000, Model Numbers 800724-2 & 800724-3 The lift motor used to raise and lower the overhead chassis will wear if not properly maintained according to the maintenance schedule. |
| Aug 19, 2016 | Class II | Z-0016-2017 | Panoramic X-ray Model PC-1000, Laser-1000 packed in a crate
Product Usage:
The PC-1000… Panoramic Corporation is performing a voluntary Medical Device Correction on certain PC-1000 X-ray systems due to the potential for a safety… |
| Jun 4, 2015 | Class II | Z-2538-2015 | Panoramic X-ray Model PC-1000
The PC-1000 will enable the user to take panoramic x-ray… Jackscrew nut part failure potentially leading to device malfunction and patient/operator injury. |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0517-2018, this FEI as the recalling firm | Jul 14, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0016-2017, this FEI as the recalling firm | Aug 19, 2016 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Oct 1, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2538-2015, this FEI as the recalling firm | Jun 4, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 28, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Panoramic Corp. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Panoramic Corp.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Panoramic Corp.
- Location
- Fort Wayne, IN (US)
- Source