REMEL, INC. FDA Inspection Record
REMEL, INC. is an FDA-registered facility (FEI 1924669) located in LENEXA, KS, US. RegDataLab tracks 7 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 3, 2023: No Action Indicated (NAI)
- Inspections on record
- 7 (2 VAI, 5 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 50, latest Oct 19, 2026
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 3, 2023 | No Action Indicated (NAI) | No citations listed |
| Oct 23, 2015 | Voluntary Action Indicated (VAI) | Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) Documentation (21 CFR 820.100(b)) Time to report - 10 days (21 CFR 806.10(b)) |
| Jul 3, 2014 | No Action Indicated (NAI) | Design plans- updated (21 CFR 820.30(b)) Investigation of device failures (21 CFR 820.198(c)) Equipment suitability & capability (21 CFR 820.72(a)) Personnel (21 CFR 820.25(a)) and 1 more |
| Jun 25, 2013 | Voluntary Action Indicated (VAI) | Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e)) Personnel requirements, Lack of or inadequate requirements (21 CFR 820.70(d)) Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b)) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) |
| Mar 7, 2013 | No Action Indicated (NAI) | No citations listed |
| Mar 18, 2010 | No Action Indicated (NAI) | Remedial action (21 CFR 820.72(b)) Procedures (21 CFR 820.100(b)) Implementing Personnel Procedures, Health, Cleanliness. (21 CFR 820.70(d)) Lack of procedures, or not maintained (21 CFR 820.40) and 2 more |
| Nov 18, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 19, 2026 | Class II | Z-0355-2027 | Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological… Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole… |
| Jun 3, 2026 | Class II | Z-2623-2026 | GC Agar Base (150mm) 10/PK R04030;
GC Agar Base (100mm) 10/PK R01460 Reduced or no recovery of some strains of Neisseria gonorrhoeae. |
| Mar 24, 2026 | Class II | Z-1806-2026 | Yeastone Broth, 11ML, 10/BOX
YY3462 Products may report incorrect AST results during quality control. Health consequences may include delayed response or the need to change to… |
| Feb 5, 2026 | Class II | Z-1474-2026 | Campy CVA Medium 100/PK, R01272 Customer complaints report low to no recovery of Campylobacter Jejuni ATCC 33291 on identified lot of medium. |
| Nov 12, 2025 | Class II | Z-0585-2026 | 1. CAMHB W/LHB, MANUAL, 11ML, 10/BOX YCP112-10
2. BRUCELLA BROTH, 11ML, 10/BOX YT3450 Products may contain contamination, which may result in a darker or brown media color. |
| Jul 11, 2025 | Class II | Z-2223-2025 | Blood Agar,5% Sheep Blood 100/PK, Product Number R01202 and 3 more products The products may contain surface and subsurface contamination of Listeria monocytogenes. |
| May 1, 2025 | Class II | Z-1947-2025 | Thermo SCIENTIFIC Sensititre Cation Adjusted AutoRead Mueller-Hinton Broth w/ TES w/ Lysed… Product may have an off color affecting perfomance |
| Feb 13, 2025 | Class II | Z-1200-2025 | remel BLOOD AGAR MacCONKEY AGAR BIPLATE, REF R02049, QTY 100, for invitro diagnostic use and 1 more product Product may contain surface and subsurface contamination of Listeria monocytogenes which is typically visible immediately upon removal from… |
| Feb 12, 2025 | Class II | Z-1395-2025 | Zebra Thermal Barcode Printer GX430t, Model no. YSW 1000 or YSW 2000, may have been sold… Power supply unit for Zebra Printers used with the device can potentially overheat or cause a fire hazard. |
| Sep 20, 2024 | Class II | Z-0266-2025 | remel BactiDrop Acridine Orange, REF 21502, 50 per package. A color change of the solution within the dropper can occur from orange-yellow to light yellow resulting in out-of-specification fluorescent… |
FDA record score
66 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
49 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0327-2023, this FEI as the recalling firm | Oct 20, 2022 | 15 x 0.6 | 9 |
| Class II recall Z-0355-2027, this FEI as the recalling firm | Oct 19, 2026 | 4 x 1 | 4 |
| Class II recall Z-2623-2026, this FEI as the recalling firm | Jun 3, 2026 | 4 x 1 | 4 |
| Class II recall Z-1806-2026, this FEI as the recalling firm | Mar 24, 2026 | 4 x 1 | 4 |
| Class II recall Z-1474-2026, this FEI as the recalling firm | Feb 5, 2026 | 4 x 1 | 4 |
| Class II recall Z-0585-2026, this FEI as the recalling firm | Nov 12, 2025 | 4 x 1 | 4 |
| Class II recall Z-2223-2025, this FEI as the recalling firm | Jul 11, 2025 | 4 x 1 | 4 |
| Class II recall Z-1947-2025, this FEI as the recalling firm | May 1, 2025 | 4 x 1 | 4 |
| Class II recall Z-1200-2025, this FEI as the recalling firm | Feb 13, 2025 | 4 x 1 | 4 |
| Class II recall Z-1395-2025, this FEI as the recalling firm | Feb 12, 2025 | 4 x 1 | 4 |
| Class II recall Z-0266-2025, this FEI as the recalling firm | Sep 20, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-2531-2023, this FEI as the recalling firm | Aug 14, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2038-2023, this FEI as the recalling firm | May 31, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0722-2022, this FEI as the recalling firm | Jan 10, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-2142-2021, this FEI as the recalling firm | Jun 18, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1561-2021, this FEI as the recalling firm | Apr 22, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1424-2021, this FEI as the recalling firm | Mar 10, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0769-2021, this FEI as the recalling firm | Dec 10, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1377-2019, this FEI as the recalling firm | Apr 17, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1400-2019, this FEI as the recalling firm | Apr 17, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-3017-2018, this FEI as the recalling firm | Aug 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0290-2018, this FEI as the recalling firm | Dec 8, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0209-2018, this FEI as the recalling firm | Nov 2, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1888-2017, this FEI as the recalling firm | Mar 24, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1400-2017, this FEI as the recalling firm | Feb 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1921-2016, this FEI as the recalling firm | Mar 24, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0443-2016, this FEI as the recalling firm | Nov 6, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 23, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2350-2015, this FEI as the recalling firm | Jun 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1841-2015, this FEI as the recalling firm | May 19, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1748-2015, this FEI as the recalling firm | May 4, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1232-2015, this FEI as the recalling firm | Feb 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0578-2015, this FEI as the recalling firm | Nov 13, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0449-2015, this FEI as the recalling firm | Oct 29, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0465-2015, this FEI as the recalling firm | Oct 23, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0200-2015, this FEI as the recalling firm | Oct 17, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2057-2014, this FEI as the recalling firm | Jun 11, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1473-2014, this FEI as the recalling firm | Mar 17, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1246-2014, this FEI as the recalling firm | Feb 18, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1112-2014, this FEI as the recalling firm | Jan 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0734-2014, this FEI as the recalling firm | Nov 21, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 25, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0537-2014, this FEI as the recalling firm | May 2, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0792-2014, this FEI as the recalling firm | Apr 9, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0530-2014, this FEI as the recalling firm | Apr 4, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0801-2013, this FEI as the recalling firm | Oct 10, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0790-2013, this FEI as the recalling firm | Aug 21, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0808-2013, this FEI as the recalling firm | Aug 21, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0806-2013, this FEI as the recalling firm | Aug 20, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing REMEL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check REMEL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Remel, Inc
- Location
- LENEXA, KS (US)
- Source
- device