CARDIOCOMMAND, INC. FDA Inspection Record
CARDIOCOMMAND, INC. is an FDA-registered facility (FEI 1927197) located in TAMPA, FL, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 30, 2026: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (2 VAI, 4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Jun 24, 2019
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 30, 2026 | Voluntary Action Indicated (VAI) | Product realization risk management processes not documented (ISO 13485:2016 Clause 7.1) |
| Dec 9, 2022 | Voluntary Action Indicated (VAI) | Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| May 17, 2017 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of Written MDR Procedures (21 CFR 803.17) |
| Jan 14, 2016 | No Action Indicated (NAI) | No citations listed |
| Feb 17, 2012 | No Action Indicated (NAI) | DHR - not or inadequately maintained (21 CFR 820.184) Documentation (21 CFR 820.100(b)) |
| Oct 24, 2008 | No Action Indicated (NAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Expired or deteriorated devices (21 CFR 820.160(a)) Results of investigation (21 CFR 820.198(e)(6)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 24, 2019 | Class III | Z-1976-2019 | TAPSYSTEM Model 2A, REF 8002A There is a potential for an assembly error resulting in the battery wire being connected to the ON/OFF power switch incorrectly. |
| May 10, 2019 | Class III | Z-1773-2019 | CardioCommand, Inc. Transesophageal Cardiac Pacing and Recording System, Model 7A, REF… Possible corrosion of the battery contact(s). |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jul 30, 2026 | 6 x 1 | 6 |
| Inspection classified VAI | Dec 9, 2022 | 6 x 0.6 | 3.6 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing CARDIOCOMMAND, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check CARDIOCOMMAND, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- CardioCommand, Inc.
- Location
- TAMPA, FL (US)
- Source
- device