Securitas Healthcare LLC FDA Inspection Record

FEI 1929691 · Inspections, Form 483s, warning letters and import refusals

Securitas Healthcare LLC is an FDA-registered facility (FEI 1929691) located in LINCOLN, NE, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jun 22, 2023: Voluntary Action Indicated (VAI)
Inspections on record
5 (3 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
13, latest Oct 12, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 22, 2023Voluntary Action Indicated (VAI)DHR - UDI not included (21 CFR 820.184(f))
Feb 6, 2017No Action Indicated (NAI)No citations listed
Nov 8, 2013Voluntary Action Indicated (VAI)Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Jan 11, 2011No Action Indicated (NAI)No citations listed
Aug 25, 2010Voluntary Action Indicated (VAI)No citations listed

Recalls

Showing the 10 most recent of 13 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Oct 12, 2026Class IIZ-0438-2027Arial Call Stations Model/Catalog Number: CSK200-900, CSK200-900MR, CSK200, CSK200MR…
Call station batteries may fully deplete without a low battery alert.
Dec 8, 2025Class IIZ-1031-2026Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3 and 1 more product
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the…
Nov 1, 2024Class IIZ-0570-2025Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the…
Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System.
Jun 20, 2024Class IIZ-2417-2024Arial Call Station, Model Numbers CSK200 and CSK200MR, is a wireless, battery-operated…
There is a risk that alarms will not be received by the Arial Wireless Emergency Call System and premature depletion of the battery.
Jun 20, 2024Class IIZ-2418-2024Arial Call Station, Model Numbers CSK200-1069 and CSK200-1069MR, is a wireless nurse call…
This is a risk that alarms will not be received by the Arial Wireless Nurse Call System and premature depletion of the battery.
Apr 11, 2024Class IIZ-1943-2024Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple…
The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all…
Sep 25, 2023Class IIZ-0220-2024Arial Water Resistant Pendant with Neck Lanyard, Models: 59360 and 2560-59360; mobile… and 2 more products
Premature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.
Feb 6, 2023Class IIZ-1250-2023Arial mobile application, model #54630 and #54640 for Android devices downloaded from the…
The mobile application for Android devices downloaded from the Google Play Store do not load per specifications and function on the Android…
Nov 17, 2021Class IIZ-0497-2022Senior Living Arial Emergency and Nurse Call Systems
Systems are not alarming due to increased memory consumption.
Jun 14, 2019Class IIZ-1996-2019STANLEY Healthcare Arial 54315 Network Manager
STANLEY Healthcare has received reports of outages on the Arial nurse call system. These outages have been traced back to the Arial¿ 54315…

FDA record score

30 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

16 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0438-2027, this FEI as the recalling firmOct 12, 20264 x 14
Class II recall Z-1031-2026, this FEI as the recalling firmDec 8, 20254 x 14
Inspection classified VAIJun 22, 20236 x 0.63.6
Class II recall Z-0570-2025, this FEI as the recalling firmNov 1, 20244 x 0.62.4
Class II recall Z-2417-2024, this FEI as the recalling firmJun 20, 20244 x 0.62.4
Class II recall Z-2418-2024, this FEI as the recalling firmJun 20, 20244 x 0.62.4
Class II recall Z-1943-2024, this FEI as the recalling firmApr 11, 20244 x 0.62.4
Class II recall Z-0220-2024, this FEI as the recalling firmSep 25, 20234 x 0.62.4
Class II recall Z-1250-2023, this FEI as the recalling firmFeb 6, 20234 x 0.62.4
Class II recall Z-0497-2022, this FEI as the recalling firmNov 17, 20214 x 0.62.4
Class II recall Z-1996-2019, this FEI as the recalling firmJun 14, 20194 x 0.251
Class II recall Z-1349-2019, this FEI as the recalling firmOct 11, 20184 x 0.251
Class II recall Z-1499-2017, this FEI as the recalling firmJun 22, 201610 years or older, so it adds no points0
Inspection classified VAINov 8, 201310 years or older, so it adds no points0
Class II recall Z-2488-2012, this FEI as the recalling firmMay 4, 201210 years or older, so it adds no points0
Inspection classified VAIAug 25, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing Securitas Healthcare LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Securitas Healthcare LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Securitas Healthcare LLC
Location
LINCOLN, NE (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.