BIOMERIEUX, INC. FDA Inspection Record

FEI 1950204 · Inspections, Form 483s, warning letters and import refusals

BIOMERIEUX, INC. is an FDA-registered facility (FEI 1950204) located in HAZELWOOD, MO, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.

Reviewing BIOMERIEUX, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check BIOMERIEUX, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

What FDA found at this site

Latest inspection
Aug 20, 2019: Voluntary Action Indicated (VAI)
Inspections on record
7 (1 OAI, 5 VAI, 1 NAI)
Last OAI inspection
Jul 30, 2014
Warning letters
1 listed by FDA for this FEI, latest Oct 10, 2014
Recalls as the recalling firm
32, latest Jul 24, 2026
Import refusals
1

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 20, 2019Voluntary Action Indicated (VAI)Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Jan 25, 2017Voluntary Action Indicated (VAI)Individual Report of Malfunction (21 CFR 803.50(a)(2))
Jun 9, 2015Voluntary Action Indicated (VAI)Sampling plans (21 CFR 820.250(b))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Jul 30, 2014Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h))
Nonconforming product control (21 CFR 820.90(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 5 more
Feb 5, 2013Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Lack of or inadequate procedures (21 CFR 820.100(a))
Design validation - software validation not performed (21 CFR 820.30(g))
and 2 more
Nov 14, 2011No Action Indicated (NAI)No citations listed
Dec 11, 2009Voluntary Action Indicated (VAI)No citations listed

Recalls

Showing the 10 most recent of 32 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 24, 2026Class IIZ-3231-2026VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference…
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a…
Jun 4, 2026Class IIZ-2715-2026VITEK 2 AST-N447 TEST KIT Product reference 424620
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch…
Jan 5, 2026Class IIZ-1236-2026VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B…
Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of…
Mar 14, 2024Class IZ-1464-2024VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility)…
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for…
Feb 6, 2023Class IIZ-1201-2023MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects…
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been…
Nov 9, 2022Class IIZ-0783-2023API LISTERIA 10 STRIPS+10 MEDIA, CATALOG 10300 and 93 more products
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Dec 20, 2021Class IIZ-0525-2022API ID 32 C Strips, Reference 32200, Rx Only, IVD API ID 32 C is a qualitative…
Risk of misidentification or delayed results with the absence of identification due to thermoforming defect.
Nov 9, 2021Class IIZ-0422-2022RAPID 20 E, REF 20701, System for the identification of Enterobacteriaceae in 4 hours, 25…
There is an error in the new version of the package insert. The Reading Table lists the incorrect color for a negative result for the CIT…
Jun 24, 2021Class IIZ-2226-2021ETEST IMIPENEM RELEBACTAM. in vitro diagnostic
Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains.
Mar 16, 2021Class IIZ-1408-2021VIDAS¿ CMV IgM 30 Tests
bioMerieux received complaints about calibration issue observed on several lots of VIDAS¿ CMV IgM (ref 30205 ; 30205-01).

Facility details

Company
Biomerieux Inc
Location
HAZELWOOD, MO (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.