BARD PERIPHERAL VASCULAR, INC. FDA Inspection Record

FEI 2020394 · Inspections, Form 483s, warning letters and import refusals

BARD PERIPHERAL VASCULAR, INC. is an FDA-registered facility (FEI 2020394) located in TEMPE, AZ, US. RegDataLab tracks 12 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Apr 21, 2026: No Action Indicated (NAI)
Inspections on record
12 (1 OAI, 1 VAI, 10 NAI)
Last OAI inspection
Jan 5, 2015
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
56, latest Aug 20, 2026
Import refusals
2

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Showing the 10 most recent of 12 inspections. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 21, 2026No Action Indicated (NAI)No citations listed
Mar 8, 2024No Action Indicated (NAI)No citations listed
Nov 2, 2018No Action Indicated (NAI)No citations listed
Feb 3, 2017No Action Indicated (NAI)No citations listed
Nov 1, 2016No Action Indicated (NAI)No citations listed
Feb 26, 2016No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Review and evaluation for investigation (21 CFR 820.198(b))
Dec 18, 2015Voluntary Action Indicated (VAI)No citations listed
Jan 5, 2015Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Conducting an Investigation (21 CFR 803.50(b)(3))
Mar 20, 2014No Action Indicated (NAI)No citations listed
Feb 24, 2014No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 56 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 20, 2026Class IIZ-3275-2026EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number…
Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code…
Dec 12, 2025Class IIZ-1367-2026Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff…
Due to damage to outer tray that can potentially compromise the sterile barrier
Nov 6, 2025Class IIZ-0887-2026Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety…
Aug 21, 2025Class IIZ-2638-2025Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal…
Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal…
Feb 5, 2025Class IZ-1244-2025Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF…
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events…
Sep 30, 2024Class IIZ-0467-2025Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF…
Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in…
Sep 4, 2024Class IIZ-0371-2025The EVSRF catheter is a sterile, single-use disposable medical device for endovenous…
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on…
Aug 2, 2024Class IIZ-3206-2024BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN…
Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the…
Jul 10, 2024Class IIZ-3062-2024Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with…
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end…
Nov 14, 2023Class IIZ-0676-2024BD Recanalization System, REF: BDRECANSYSTEM
Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and…

FDA record score

86 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

58 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-1244-2025, this FEI as the recalling firmFeb 5, 202515 x 115
Class II recall Z-3275-2026, this FEI as the recalling firmAug 20, 20264 x 14
Class II recall Z-1367-2026, this FEI as the recalling firmDec 12, 20254 x 14
Class II recall Z-0887-2026, this FEI as the recalling firmNov 6, 20254 x 14
Class II recall Z-2638-2025, this FEI as the recalling firmAug 21, 20254 x 14
Class II recall Z-0467-2025, this FEI as the recalling firmSep 30, 20244 x 0.62.4
Class II recall Z-0371-2025, this FEI as the recalling firmSep 4, 20244 x 0.62.4
Class II recall Z-3206-2024, this FEI as the recalling firmAug 2, 20244 x 0.62.4
Class II recall Z-3062-2024, this FEI as the recalling firmJul 10, 20244 x 0.62.4
Class II recall Z-0676-2024, this FEI as the recalling firmNov 14, 20234 x 0.62.4
Class II recall Z-0643-2024, this FEI as the recalling firmNov 8, 20234 x 0.62.4
Class II recall Z-0027-2024, this FEI as the recalling firmAug 10, 20234 x 0.62.4
Class II recall Z-1564-2023, this FEI as the recalling firmApr 11, 20234 x 0.62.4
Class II recall Z-1260-2023, this FEI as the recalling firmFeb 28, 20234 x 0.62.4
Class II recall Z-1051-2023, this FEI as the recalling firmJan 4, 20234 x 0.62.4
Class II recall Z-1108-2023, this FEI as the recalling firmDec 22, 20224 x 0.62.4
Class II recall Z-1273-2022, this FEI as the recalling firmApr 22, 20224 x 0.62.4
Class II recall Z-0790-2022, this FEI as the recalling firmJan 20, 20224 x 0.62.4
Class II recall Z-0153-2022, this FEI as the recalling firmAug 24, 20214 x 0.251
Class II recall Z-0002-2022, this FEI as the recalling firmJul 27, 20214 x 0.251
Class II recall Z-2247-2021, this FEI as the recalling firmJun 23, 20214 x 0.251
Class II recall Z-1550-2021, this FEI as the recalling firmMar 25, 20214 x 0.251
Class II recall Z-1289-2021, this FEI as the recalling firmJan 14, 20214 x 0.251
Class II recall Z-1897-2021, this FEI as the recalling firmDec 14, 20204 x 0.251
Class II recall Z-0080-2021, this FEI as the recalling firmOct 1, 20204 x 0.251
Class II recall Z-1689-2020, this FEI as the recalling firmMar 17, 20204 x 0.251
Class II recall Z-1964-2020, this FEI as the recalling firmJan 15, 20204 x 0.251
Class II recall Z-1431-2020, this FEI as the recalling firmJan 3, 20204 x 0.251
Class II recall Z-0724-2020, this FEI as the recalling firmOct 25, 20194 x 0.251
Class II recall Z-1570-2020, this FEI as the recalling firmOct 1, 20194 x 0.251
Class II recall Z-0917-2020, this FEI as the recalling firmAug 14, 20194 x 0.251
Class II recall Z-2208-2020, this FEI as the recalling firmJun 14, 20194 x 0.251
Class II recall Z-2368-2019, this FEI as the recalling firmMay 31, 20194 x 0.251
Class II recall Z-2209-2019, this FEI as the recalling firmMay 2, 20194 x 0.251
Class II recall Z-1637-2020, this FEI as the recalling firmApr 12, 20194 x 0.251
Class II recall Z-0718-2020, this FEI as the recalling firmFeb 18, 20194 x 0.251
Class II recall Z-0582-2020, this FEI as the recalling firmFeb 14, 20194 x 0.251
Class II recall Z-2080-2019, this FEI as the recalling firmJun 25, 20184 x 0.251
Class II recall Z-1802-2018, this FEI as the recalling firmApr 16, 20184 x 0.251
Class II recall Z-1760-2018, this FEI as the recalling firmFeb 8, 20184 x 0.251
Class II recall Z-0044-2018, this FEI as the recalling firmSep 22, 20174 x 0.251
Class II recall Z-1900-2017, this FEI as the recalling firmMar 28, 20174 x 0.251
Class II recall Z-2547-2016, this FEI as the recalling firmJul 6, 201610 years or older, so it adds no points0
Class II recall Z-1068-2016, this FEI as the recalling firmFeb 9, 201610 years or older, so it adds no points0
Class II recall Z-1396-2016, this FEI as the recalling firmFeb 4, 201610 years or older, so it adds no points0
Class II recall Z-0785-2016, this FEI as the recalling firmJan 5, 201610 years or older, so it adds no points0
Inspection classified VAIDec 18, 201510 years or older, so it adds no points0
Class II recall Z-0223-2016, this FEI as the recalling firmOct 19, 201510 years or older, so it adds no points0
Class II recall Z-1586-2015, this FEI as the recalling firmApr 9, 201510 years or older, so it adds no points0
Class II recall Z-1413-2015, this FEI as the recalling firmMar 23, 201510 years or older, so it adds no points0
Class II recall Z-1447-2015, this FEI as the recalling firmMar 17, 201510 years or older, so it adds no points0
Class II recall Z-0737-2016, this FEI as the recalling firmMar 13, 201510 years or older, so it adds no points0
Inspection classified OAIJan 5, 201510 years or older, so it adds no points0
Class II recall Z-2648-2014, this FEI as the recalling firmAug 26, 201410 years or older, so it adds no points0
Class I recall Z-0007-2014, this FEI as the recalling firmSep 30, 201310 years or older, so it adds no points0
Class II recall Z-1533-2013, this FEI as the recalling firmMay 28, 201310 years or older, so it adds no points0
Class II recall Z-1524-2013, this FEI as the recalling firmApr 16, 201310 years or older, so it adds no points0
Class II recall Z-1526-2013, this FEI as the recalling firmJan 23, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing BARD PERIPHERAL VASCULAR, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check BARD PERIPHERAL VASCULAR, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Bard Peripheral Vascular Inc
Location
TEMPE, AZ (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.