MICROAIRE SURGICAL INSTRUMENTS LLC FDA Inspection Record
MICROAIRE SURGICAL INSTRUMENTS LLC is an FDA-registered facility (FEI 2020601) located in CHARLOTTESVILLE, VA, US. RegDataLab tracks 3 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Mar 17, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (1 OAI, 2 VAI)
- Last OAI inspection
- Aug 22, 2014
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Dec 29, 2014. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 12, latest May 7, 2025
- Import refusals
- 6
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 17, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) Reporting Information Reasonably Known (21 CFR 803.50(b)(1)) |
| Aug 22, 2014 | Official Action Indicated (OAI) | Design review - Lack of or inadequate procedures (21 CFR 820.30(e)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 5 more |
| Mar 31, 2011 | Voluntary Action Indicated (VAI) | Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Equipment control activity documentation (21 CFR 820.72(a)) Equipment Installation, Placement, Specified Requirements (21 CFR 820.70(g)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
Warning letters
- Dec 29, 2014: Warning letter listed on FDA's Data Dashboard. View compliance record (440739).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 7, 2025 | Class II | Z-1975-2025 | Brand Name: SmartRelease ESTR (Endoscopic soft tissue release
system), Onyx Blade… Potential that the blade can unintentionally cut tissue prior to the user deploying the blade. |
| Jun 26, 2019 | Class II | Z-2222-2019 | SmartRelease Endoscopic Carpal Tunnel Release Handpiece (ECTR handpiece), part numbers… There was a higher than normal incidence rate of the blade not retracting when tightening. |
| Apr 19, 2018 | Class II | Z-2107-2018 | MicroAire SmartRelease Endoscopic System, Endoscopic Carpal Tunnel Release, SmartRelease… The Instructions for Use (IFU) is incorrect with regards to sterilization parameters and must be replaced. |
| May 22, 2017 | Class II | Z-0412-2018 | SST TWIST DRILL, STANDARD, 1.6mm Dia. x 127mm, REF 8054-010, STERILE
This product line… One lot was found to contain both the correct twist drill (1.6mm) as well as an incorrect twist drill (2.0mm) |
| Feb 21, 2017 | Class II | Z-1407-2017 | MicroAire K-Wires Mislabeling on three lots of K-Wires. The incorrect product description " dual trocar" instead of "single trocar" was on the product label… |
| May 6, 2016 | Class II | Z-2294-2016 | Endotine Transbleph 3.5 (CFD-080-0167) MicroAire has initiated a limited recall (29 Units)on the Endotine Transbleph 3.5 (CFD-080-0167), Lot # 285373, due to minor mislabeling (… |
| Mar 11, 2016 | Class II | Z-1834-2016 | Pneumatic Impactor, General Surgery, With Microaire Hose Connector
Professional Use… MicroAire discovered that 20 suspect Impactor instruments ( MicroAire Pneumatic Impactor) were shipped to end users and were missing the… |
| Feb 8, 2016 | Class II | Z-1323-2016 | MicroAire PAL-R4011XL - 40cm x 4 mm Del Vecchio French Rasp aspiration cannula Complaints that the cannula bent/broke during surgery. |
| Feb 8, 2016 | Class II | Z-1328-2016 | MicroAire PAL¿ Cannula
Product Usage - Power Aspiration Device MicroAire has initiated a recall on product PAL-R4011XL (40cm x 4mm Del Vecchio French Rasp cannula), due to our receipt of three (3)… |
| Jun 12, 2015 | Class II | Z-2821-2015 | ASP-1020 LipoTower with Aspiration Pump. Medical use. and 1 more product There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails. |
FDA record score
8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
16 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1975-2025, this FEI as the recalling firm | May 7, 2025 | 4 x 1 | 4 |
| Class II recall Z-2222-2019, this FEI as the recalling firm | Jun 26, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2107-2018, this FEI as the recalling firm | Apr 19, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0412-2018, this FEI as the recalling firm | May 22, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1407-2017, this FEI as the recalling firm | Feb 21, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2294-2016, this FEI as the recalling firm | May 6, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 17, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1834-2016, this FEI as the recalling firm | Mar 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1323-2016, this FEI as the recalling firm | Feb 8, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1328-2016, this FEI as the recalling firm | Feb 8, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2821-2015, this FEI as the recalling firm | Jun 12, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Dec 29, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 22, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2158-2013, this FEI as the recalling firm | Aug 12, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1792-2012, this FEI as the recalling firm | Apr 26, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 31, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing MICROAIRE SURGICAL INSTRUMENTS LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check MICROAIRE SURGICAL INSTRUMENTS LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- MicroAire Surgical Instruments, LLC
- Location
- CHARLOTTESVILLE, VA (US)
- Source
- device