HARDY DIAGNOSTICS FDA Inspection Record
HARDY DIAGNOSTICS is an FDA-registered facility (FEI 2022807) located in Orcutt, CA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jun 9, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (1 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 14, latest Jun 8, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 9, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Documentation (21 CFR 820.100(b)) Training records (21 CFR 820.25(b)) |
| Nov 28, 2018 | No Action Indicated (NAI) | No citations listed |
| Nov 5, 2015 | No Action Indicated (NAI) | No citations listed |
| Apr 15, 2011 | No Action Indicated (NAI) | Quality audits - auditor independence (21 CFR 820.22) Buildings (21 CFR 820.70(f)) Lack of or inadequate procedures for handling (21 CFR 820.140) QSR (21 CFR 820.186) and 2 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 8, 2026 | Class II | Z-2887-2026 | HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435
HardyDisk AST Cefiderocol 30¿g… Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility… |
| May 6, 2025 | Class II | Z-2041-2025 | HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility… Identified lots failed incoming quality control testing. Lots exhibited breakthrough growth of Klebsiella pneumoniae. |
| Apr 16, 2021 | Class II | Z-1835-2021 | Quickslide GramPro 1 Automated Gram Stanier is a microscope slide stainer. Instruments were assembled and serviced using a defective lot of tubing, which is part of the pump mechanism within the unit. |
| Dec 21, 2020 | Class II | Z-0919-2021 | Brain Heart Infusion Agar with 6 ug/ml Vancomycin Cat no: G14 Lot no: 472698 Manufactured… Complaints were received reporting of Staphylococcus aureus (ATCC 29213) breakthrough when the strain should have been inhibited. |
| Oct 23, 2020 | Class II | Z-0655-2021 | Viral Transport Medium, 3ml, Ref: R99, Container type: Tube, Packaged: 20 tubes/box, IVD… There is a potential for contamination within Viral Transport Media. Use of contaminated Viral Transport Media for the transport of patient… |
| Oct 6, 2020 | Class III | Z-0441-2021 | Strep B Carrot Broth One-Step, Ref. Z40, UDI: 00819398026499 Streptococcus culture medium contains an incorrect expiration date. The culture medium outer box labeling contains an expiration date of… |
| Nov 14, 2019 | Class II | Z-2298-2020 | HardyCHROM MRSA,REF; G307, LOT: 447694, A selective and differential chromogenic medium… False Positive results due to the MSSA (methicillin-susceptible) strain Staphylococcus aureus (ATCC 29213) breakthrough when this strain… |
| May 24, 2016 | Class II | Z-2146-2016 | QuickSlide MGS-80, Automated Mutli-Slide Gram Stainer/GramPRO-80.
Systematically… Discrepant culture and gram stain results, which could potentially lead to misdiagnosis or improper treatment of a patient. |
| Mar 21, 2016 | Class II | Z-1571-2016 | Hardy Diagnostics GBS Detect
Cat no. A300.
Used for the isolation and detection by… Performance failure; poor hemolytic reaction of non or weakly hemolytic group B streptococcus (GBS) with the target QC organism… |
| Nov 18, 2015 | Class II | Z-0470-2016 | FB Broth, 10ml
(Fastidious Bacteria Broth)
Cat. no: K31 Lot: 15231
A fastidious… The firm is recalling FB Broth Cat. no. K31, Lot no. 15231 because of potential Burkholderia fungorum contamination. |
FDA record score
16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
14 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2887-2026, this FEI as the recalling firm | Jun 8, 2026 | 4 x 1 | 4 |
| Class II recall Z-2041-2025, this FEI as the recalling firm | May 6, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Jun 9, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-1835-2021, this FEI as the recalling firm | Apr 16, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0919-2021, this FEI as the recalling firm | Dec 21, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0655-2021, this FEI as the recalling firm | Oct 23, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2298-2020, this FEI as the recalling firm | Nov 14, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2146-2016, this FEI as the recalling firm | May 24, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1571-2016, this FEI as the recalling firm | Mar 21, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0470-2016, this FEI as the recalling firm | Nov 18, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2463-2015, this FEI as the recalling firm | Jul 9, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2273-2015, this FEI as the recalling firm | Jun 26, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2192-2015, this FEI as the recalling firm | Jun 24, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1444-2015, this FEI as the recalling firm | Mar 4, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing HARDY DIAGNOSTICS as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check HARDY DIAGNOSTICS yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Hardy Diagnostics
- Location
- Orcutt, CA (US)
- Source
- device