TALLADIUM, INC. FDA Inspection Record
TALLADIUM, INC. is an FDA-registered facility (FEI 2023129) located in Santa Clarita, CA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 18, 2015: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Oct 9, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 18, 2015 | Voluntary Action Indicated (VAI) | Design review - documentation (21 CFR 820.30(e)) Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b)) Lack of or inadequate DHR procedures (21 CFR 820.184) |
| Feb 9, 2012 | No Action Indicated (NAI) | No citations listed |
| Dec 11, 2009 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 9, 2020 | Class III | Z-0512-2021 | Luminesse ML PMMA, 98x20mm, REF: 5336-B1 - Product Usage: used as a milling blank in the… Heat-cured polymethylmethacrylate (PMMA) milling blank product used to make long-term temporary crowns and bridge restorations has incorrect… |
| Aug 6, 2020 | Class II | Z-2943-2020 | Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0… The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was… |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2943-2020, this FEI as the recalling firm | Aug 6, 2020 | 4 x 0.25 | 1 |
| Inspection classified VAI | Aug 18, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 11, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing TALLADIUM, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.