STAAR SURGICAL COMPANY FDA Inspection Record

FEI 2023826 · Inspections, Form 483s, warning letters and import refusals

STAAR SURGICAL COMPANY is an FDA-registered facility (FEI 2023826) located in MONROVIA, CA, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
May 18, 2018: Voluntary Action Indicated (VAI)
Inspections on record
7 (2 OAI, 3 VAI, 2 NAI)
Last OAI inspection
Feb 4, 2015
Warning letters
1 listed by FDA for this FEI, latest May 21, 2014
Recalls as the recalling firm
3, latest Sep 29, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
May 18, 2018Voluntary Action Indicated (VAI)No citations listed
Feb 4, 2015Official Action Indicated (OAI)No citations listed
Feb 4, 2015No Action Indicated (NAI)Incorrect translation to production specifications (21 CFR 820.30(h))
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Software validation for automated processes (21 CFR 820.70(i))
and 6 more
Mar 21, 2014Official Action Indicated (OAI)No citations listed
Sep 4, 2013Voluntary Action Indicated (VAI)No citations listed
Jan 25, 2012No Action Indicated (NAI)No citations listed
Mar 4, 2009Voluntary Action Indicated (VAI)Complete files maintained (21 CFR 820.198(a))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Analysis of data/reports from data sources (21 CFR 820.100(a)(1))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Sep 29, 2023Class IIZ-0464-2024EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens. and 10 more products
Their is a potential that intraocular implant devices may not meet specifications.
May 2, 2016Class IIZ-1964-2016STAAR AQ Cartridge-FP, guide and delivery system for Intraocular lenses.
The STAAR AQ Cartridge-FP may malfunction and may crack during loading or delivery of the IOL (Intraocular Lenses).
Jun 10, 2013Class IIZ-1746-2013nanoFLEX CC4204A Intraocular Lens Correct aphakia in persons 60 years of age or older… and 1 more product
STAAR Surgical Company is recalling a limited number of nanoFLEX CC4204A and Afinity CQ2015A intraocular lenses due to the potential…

FDA record score

4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0464-2024, this FEI as the recalling firmSep 29, 20234 x 0.62.4
Inspection classified VAIMay 18, 20186 x 0.251.5
Class II recall Z-1964-2016, this FEI as the recalling firmMay 2, 201610 years or older, so it adds no points0
Inspection classified OAIFeb 4, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsMay 21, 201410 years or older, so it adds no points0
Inspection classified OAIMar 21, 201410 years or older, so it adds no points0
Inspection classified VAISep 4, 201310 years or older, so it adds no points0
Class II recall Z-1746-2013, this FEI as the recalling firmJun 10, 201310 years or older, so it adds no points0
Inspection classified VAIMar 4, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing STAAR SURGICAL COMPANY as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.