QUIDEL CORP. FDA Inspection Record
QUIDEL CORP. is an FDA-registered facility (FEI 2024674) located in SAN DIEGO, CA, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Apr 10, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 9, latest Apr 6, 2026
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 10, 2017 | Voluntary Action Indicated (VAI) | No citations listed |
| Jan 13, 2014 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 6, 2026 | Class II | Z-2169-2026 | QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. Product has the potential for false positive results |
| Jun 17, 2025 | Class II | Z-2545-2025 | QuickVue: Dipstick Strep A Test: 50T, REF: 20108; 25T, REF: 20125; 50T, SCN, REF: 20108SC Dipstick strep A test has potential for false positive results. |
| Mar 5, 2024 | Class II | Z-1636-2024 | QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20-U, 20 Tests and 1 more product There are two windows on the test device, the control line window and the test results window. The line for the test results window were… |
| Sep 25, 2020 | Class II | Z-0428-2021 | QUIDEL Sofia SARS Antigen FIA IVD REF 20374 Due to false positive results. |
| Sep 17, 2020 | Class II | Z-0124-2021 | Sofia SARS Antigen FIA Package Insert
The Sofia SARS Antigen FIA is a lateral flow… Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport… |
| Dec 15, 2017 | Class II | Z-1183-2018 | Sofia 2 Analyzer Catalog Number 20299
Sofia 2 is a bench top analyzer intended to be… The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when… |
| Jun 22, 2017 | Class II | Z-2803-2017 | InflammaDry(R) AN RPS DIAGNOSTIC SOLUTION, Sample Collectors
InflammaDry(R) is a rapid… In-house testing revealed the presence of plastic spurs at the tips of some Sample Collectors. |
| Jan 25, 2017 | Class II | Z-1231-2017 | Influenza Transport Swab Packages, Reference #20171,
One lot with 1010 packages (each… Quidel Corporation is recalling Influenza Transport Swab Packages because it may include an incorrect sample swab mixed in with the proper… |
| Dec 3, 2012 | Class II | Z-0819-2013 | Quidel, Sofia Influenza A + B FIA, Kit #20218 (25-Test Kit), in vitro diagnostic. Quidel Corporation has initiated a voluntary recall of specific lots of the Sofia Influenza A+B FIA kits because of false positive results… |
FDA record score
17 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2169-2026, this FEI as the recalling firm | Apr 6, 2026 | 4 x 1 | 4 |
| Class II recall Z-2545-2025, this FEI as the recalling firm | Jun 17, 2025 | 4 x 1 | 4 |
| Class II recall Z-1636-2024, this FEI as the recalling firm | Mar 5, 2024 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Apr 10, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0428-2021, this FEI as the recalling firm | Sep 25, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0124-2021, this FEI as the recalling firm | Sep 17, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1183-2018, this FEI as the recalling firm | Dec 15, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2803-2017, this FEI as the recalling firm | Jun 22, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1231-2017, this FEI as the recalling firm | Jan 25, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0819-2013, this FEI as the recalling firm | Dec 3, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |