Lacrimedics Inc FDA Inspection Record

FEI 2024818 · Inspections, Form 483s, warning letters and import refusals

Lacrimedics Inc is an FDA-inspected facility (FEI 2024818) located in Dupont, WA, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jan 23, 2020: Voluntary Action Indicated (VAI)
Inspections on record
5 (1 VAI, 4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
2, latest Jan 17, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jan 23, 2020Voluntary Action Indicated (VAI)Procedures not adequately established or maintained (21 CFR 820.40)
Lack of or inadequate procedures (21 CFR 820.100(a))
Documented evaluation (21 CFR 820.50(a)(1))
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))
Mar 15, 2016No Action Indicated (NAI)DHR - not or inadequately maintained (21 CFR 820.184)
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design input - documentation (21 CFR 820.30(c))
Equipment design and installation (21 CFR 820.70(g))
Apr 26, 2013No Action Indicated (NAI)No citations listed
Jun 23, 2011No Action Indicated (NAI)No citations listed
Jun 9, 2009No Action Indicated (NAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 17, 2022Class IIZ-0708-2022OPAQUE Herrick Lacrimal Plugs, HP3, 0.3 mm, HP5, 0.5mm, HP7, 0.7.mm, STERILE, EO, Rx only… and 1 more product
The sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
Nov 26, 2014Class IIIZ-1081-2015VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs…
Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4…

FDA record score

4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

2 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0708-2022, this FEI as the recalling firmJan 17, 20224 x 0.62.4
Inspection classified VAIJan 23, 20206 x 0.251.5

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Lacrimedics Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Lacrimedics Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Lacrimedics Inc
Location
Dupont, WA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.