MICROBIOLOGICS, INC. FDA Inspection Record

FEI 2150138 · Inspections, Form 483s, warning letters and import refusals

MICROBIOLOGICS, INC. is an FDA-registered facility (FEI 2150138) located in Saint Cloud, MN, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 1, 2021: Voluntary Action Indicated (VAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
43, latest Dec 2, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 1, 2021Voluntary Action Indicated (VAI)Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2))
Nov 26, 2019Voluntary Action Indicated (VAI)Documentation (21 CFR 820.100(b))
Limitations of procedure (21 CFR 809.10(b)(10))
Feb 6, 2017No Action Indicated (NAI)No citations listed
Jul 18, 2013No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 43 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 2, 2025Class IIZ-0984-2026KWIK-STIK 2-Pack, Campylobacter jejuni subsp. jejuni derived from ATCC 29428, Catalog… and 2 more products
Affected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
Nov 13, 2025Class IIZ-0876-2026KWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from…
Some products packaged for lot 818-111-7 may have been packaged in pouches labeled as lot 857-53-10.
Jun 27, 2025Class IIZ-2194-2025LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF… and 1 more product
KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76…
Jun 17, 2025Class IIZ-2124-2025KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L
Affected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality…
May 27, 2025Class IIZ-1957-2025KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella…
Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was…
Feb 20, 2025Class IIZ-1371-2025Helix Elite" Inactivated Standard: Inactivated SARS-CoV-2 Whole Virus (Pellet)…
The A549 human cell target (human cells for sample adequacy control) may give a late Ct value.
Feb 20, 2025Class IIZ-1454-2025Helix Elite Inactivated Standard: Inactivated Influenza A/B and Respiratory Syncytial…
The RSV target may give a late Ct value and could potentially not pass QC.
Jan 20, 2025Class IIZ-1163-2025Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet)… and 1 more product
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the…
Mar 6, 2024Class IIZ-1483-2024KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K…
The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus…
Jan 8, 2024Class IIZ-1075-2024Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive…

FDA record score

67 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

34 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0984-2026, this FEI as the recalling firmDec 2, 20254 x 14
Class II recall Z-0876-2026, this FEI as the recalling firmNov 13, 20254 x 14
Class II recall Z-2194-2025, this FEI as the recalling firmJun 27, 20254 x 14
Class II recall Z-2124-2025, this FEI as the recalling firmJun 17, 20254 x 14
Class II recall Z-1957-2025, this FEI as the recalling firmMay 27, 20254 x 14
Class II recall Z-1371-2025, this FEI as the recalling firmFeb 20, 20254 x 14
Class II recall Z-1454-2025, this FEI as the recalling firmFeb 20, 20254 x 14
Class II recall Z-1163-2025, this FEI as the recalling firmJan 20, 20254 x 14
Inspection classified VAIDec 1, 20216 x 0.63.6
Class II recall Z-1483-2024, this FEI as the recalling firmMar 6, 20244 x 0.62.4
Class II recall Z-1075-2024, this FEI as the recalling firmJan 8, 20244 x 0.62.4
Class II recall Z-0307-2024, this FEI as the recalling firmOct 18, 20234 x 0.62.4
Class II recall Z-0954-2022, this FEI as the recalling firmMar 24, 20224 x 0.62.4
Class II recall Z-0955-2022, this FEI as the recalling firmMar 22, 20224 x 0.62.4
Class II recall Z-0844-2022, this FEI as the recalling firmMar 3, 20224 x 0.62.4
Class II recall Z-0784-2022, this FEI as the recalling firmFeb 25, 20224 x 0.62.4
Inspection classified VAINov 26, 20196 x 0.251.5
Class II recall Z-2486-2021, this FEI as the recalling firmAug 20, 20214 x 0.251
Class II recall Z-2487-2021, this FEI as the recalling firmAug 20, 20214 x 0.251
Class II recall Z-0511-2022, this FEI as the recalling firmDec 16, 20204 x 0.251
Class II recall Z-0302-2021, this FEI as the recalling firmOct 8, 20204 x 0.251
Class II recall Z-0402-2021, this FEI as the recalling firmOct 8, 20204 x 0.251
Class II recall Z-0582-2022, this FEI as the recalling firmOct 1, 20204 x 0.251
Class II recall Z-0585-2022, this FEI as the recalling firmFeb 18, 20204 x 0.251
Class II recall Z-1690-2020, this FEI as the recalling firmFeb 12, 20204 x 0.251
Class II recall Z-1096-2020, this FEI as the recalling firmJan 3, 20204 x 0.251
Class II recall Z-0812-2020, this FEI as the recalling firmDec 16, 20194 x 0.251
Class II recall Z-1730-2019, this FEI as the recalling firmApr 9, 20194 x 0.251
Class II recall Z-0982-2019, this FEI as the recalling firmJan 31, 20194 x 0.251
Class II recall Z-0697-2022, this FEI as the recalling firmJun 6, 20174 x 0.251
Class II recall Z-0703-2022, this FEI as the recalling firmDec 1, 201510 years or older, so it adds no points0
Class II recall Z-0712-2022, this FEI as the recalling firmOct 6, 201410 years or older, so it adds no points0
Class II recall Z-1021-2013, this FEI as the recalling firmFeb 11, 201310 years or older, so it adds no points0
Class II recall Z-1936-2012, this FEI as the recalling firmMay 29, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing MICROBIOLOGICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check MICROBIOLOGICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Microbiologics Inc
Location
Saint Cloud, MN (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.