ORTHOFIX US LLC FDA Inspection Record
ORTHOFIX US LLC is an FDA-registered facility (FEI 2183449) located in Lewisville, TX, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 17, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (2 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 16, latest Jun 10, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 17, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) |
| Jan 3, 2014 | No Action Indicated (NAI) | No citations listed |
| Jun 24, 2011 | Voluntary Action Indicated (VAI) | Design validation - user needs and intended uses (21 CFR 820.30(g)) |
| Feb 23, 2010 | No Action Indicated (NAI) | No citations listed |
| Dec 16, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 10, 2026 | Class II | Z-2729-2026 | Orthofix MINIRAIL, STANDARD LENGTHENER
Model/Catalog Number: M101, Unilateral… and 3 more products Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and… |
| Nov 3, 2025 | Class III | Z-0891-2026 | Brand Name: CONSTRUX" Mini PTC Spacer System
Product Name: CONSTRUX" Mini PTC Spacer… and 5 more products Labeling contains claims that are not consistently present. |
| Sep 19, 2025 | Class II | Z-0478-2026 | Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The… Polyaxial screws do not meet performance standards. |
| Jun 24, 2025 | Class II | Z-2555-2025 | PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label… |
| Feb 20, 2025 | Class II | Z-1432-2025 | Orthofix Pillar SA Ti Spacer System Intervertebral Body Fusion Spinal Device, sterile… The product is mislabeled with the incorrect anterior height of 10mm, but the laser marking on the implant and corresponding trial both show… |
| Apr 6, 2018 | Class II | Z-2116-2018 | Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only… and 19 more products Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a… |
| Feb 23, 2018 | Class II | Z-1363-2018 | ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003 and 1 more product Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and… |
| Aug 2, 2017 | Class II | Z-3127-2017 | Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile… The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or… |
| Apr 27, 2016 | Class II | Z-1881-2016 | Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a… There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is… |
| Jan 8, 2016 | Class II | Z-0797-2016 | Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512)
Intended for… There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst… |
FDA record score
23 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
17 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2729-2026, this FEI as the recalling firm | Jun 10, 2026 | 4 x 1 | 4 |
| Class II recall Z-0478-2026, this FEI as the recalling firm | Sep 19, 2025 | 4 x 1 | 4 |
| Class II recall Z-2555-2025, this FEI as the recalling firm | Jun 24, 2025 | 4 x 1 | 4 |
| Class II recall Z-1432-2025, this FEI as the recalling firm | Feb 20, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Mar 17, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-2116-2018, this FEI as the recalling firm | Apr 6, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1363-2018, this FEI as the recalling firm | Feb 23, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-3127-2017, this FEI as the recalling firm | Aug 2, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1881-2016, this FEI as the recalling firm | Apr 27, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0797-2016, this FEI as the recalling firm | Jan 8, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1749-2015, this FEI as the recalling firm | May 18, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1819-2015, this FEI as the recalling firm | May 11, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0855-2014, this FEI as the recalling firm | Nov 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1089-2014, this FEI as the recalling firm | Oct 29, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0896-2013, this FEI as the recalling firm | Feb 1, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0895-2013, this FEI as the recalling firm | Aug 6, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 24, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing ORTHOFIX US LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ORTHOFIX US LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Orthofix U.S. LLC
- Location
- Lewisville, TX (US)
- Source
- device