Welch Allyn Inc (Baxter Affiliate) FDA Inspection Record
Welch Allyn Inc (Baxter Affiliate) is an FDA-inspected facility (FEI 2183461) located in Milwaukee, WI, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 15, 2022: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (4 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 6, latest Sep 3, 2021
- Import refusals
- 10 or more
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 15, 2022 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) |
| Apr 7, 2017 | Voluntary Action Indicated (VAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Software validation for automated processes (21 CFR 820.70(i)) Documentation (21 CFR 820.80(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 3 more |
| Sep 18, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation - software validation documentation (21 CFR 820.30(g)) |
| May 10, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Review and evaluation for investigation (21 CFR 820.198(b)) Investigation of device failures (21 CFR 820.198(c)) and 3 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 3, 2021 | Class II | Z-0011-2022 | ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and… The devices malfunction under specific operator workflows. |
| Mar 31, 2021 | Class II | Z-1482-2021 | ELI 280 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software… and 3 more products Devices do not meet IEC 60601-2-27 requirements as labeled. |
| Nov 11, 2020 | Class II | Z-0553-2021 | ELI 380 Electrocardiograph - Product Usage: intended to be a high-performance… The radio within the device can become disassociated with the wireless access point. |
| Mar 18, 2020 | Class II | Z-1810-2020 | Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI… and 9 more products Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment. |
| May 29, 2015 | Class II | Z-1796-2015 | ELI 380 Electrocardiograph.
The ELI 380 is intended to be a high-performance… Mortara Instrument, Inc. has recently become aware of a potential safety hazard involving our ELI 380 electrocardiograph. When used in a… |
| Oct 7, 2014 | Class II | Z-0110-2015 | Mortara Surveyor S12 and Surveyor S19 Patient Monitors.
Product Usage:
They are… Martara Instrument, Inc. has recently became aware of a potential safety hazard involving our Surveyor rand S12 S19 Patient Monitors. A… |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jul 15, 2022 | 6 x 0.6 | 3.6 |
| Inspection classified VAI | Apr 7, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0011-2022, this FEI as the recalling firm | Sep 3, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1482-2021, this FEI as the recalling firm | Mar 31, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0553-2021, this FEI as the recalling firm | Nov 11, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1810-2020, this FEI as the recalling firm | Mar 18, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1796-2015, this FEI as the recalling firm | May 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0110-2015, this FEI as the recalling firm | Oct 7, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 18, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 10, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 29, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 29, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 29, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
Reviewing Welch Allyn Inc (Baxter Affiliate) as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Welch Allyn Inc (Baxter Affiliate): recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Welch Allyn Inc (Baxter Affiliate) → Check this supplier
Facility details
- FDA Legal Name
- Welch Allyn Inc (Baxter affiliate)
- Company
- Welch Allyn Inc (Baxter Affiliate)
- Location
- Milwaukee, WI (US)
- Source
- inspections