Immucor GTI Diagnostics, Inc. FDA Inspection Record
Immucor GTI Diagnostics, Inc. is an FDA-registered facility (FEI 2183608) located in Waukesha, WI, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 1, 2016: No Action Indicated (NAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Jun 28, 2018
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 1, 2016 | No Action Indicated (NAI) | Design plans- updated (21 CFR 820.30(b)) Design input - documentation (21 CFR 820.30(c)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Nonconforming product control (21 CFR 820.90(a)) and 3 more |
| Jan 13, 2010 | Voluntary Action Indicated (VAI) | Acceptance criteria and essential outputs (21 CFR 820.30(d)) Design review - Lack of or inadequate procedures (21 CFR 820.30(e)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Documentation (21 CFR 820.100(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 28, 2018 | Class III | B-0695-2018 | ThromboType, Product No. 303481, UDI 10888234500285 Immucor GTI Diagnostics ThromboType kit, with the potential for poor amplification for the HPA-1a and HPA-3a reactions, was distributed. |
| Jan 31, 2018 | Class III | B-0347-2018 | LIFECODES HLA-A SSO Typing Kit (that utilize MATCHIT! DNA Software v1.2); Product No… and 1 more product Immucor GTI Diagnostics' LIFECODES HLA-A SSO Typing Kits and LIFECODES HLA-A eRES SSO Typing Kits that utilize MATCH IT! DNA Software v1.2… |
| Sep 21, 2017 | Class II | B-0001-2018 | LIFECODES ThromboType 1 Kit, Product No. 303479, UDI No. 10888234500261 LIFECODES ThromboType 1 Kit, with the potential to generate a failed assay or sample mistype, was distributed. |
| Dec 15, 2014 | Class II | Z-1001-2015 | PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits: 1… Some Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead to slightly increased Optical Density (OD)… |
| Feb 4, 2013 | Class II | Z-1022-2013 | Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45
GTI PF4 Enhanced is… Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit… |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall B-0001-2018, this FEI as the recalling firm | Sep 21, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1001-2015, this FEI as the recalling firm | Dec 15, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1022-2013, this FEI as the recalling firm | Feb 4, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 13, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Immucor GTI Diagnostics, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Immucor GTI Diagnostics, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Immucor GTI Diagnostics, Inc
- Location
- Waukesha, WI (US)
- Source