Vital Signs Devices, A GE Healthcare Company FDA Inspection Record
Vital Signs Devices, A GE Healthcare Company is an FDA-inspected facility (FEI 2242551) located in Totowa, NJ, US. RegDataLab tracks 1 FDA inspection and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 15, 2011: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 listed by FDA for this FEI, latest Dec 1, 2008
- Recalls as the recalling firm
- 3, latest Jul 18, 2013
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 15, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Equipment control activity documentation (21 CFR 820.72(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) Submission of Supplement to Individual Report Within One Month (21 CFR 803.56) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 18, 2013 | Class II | Z-1959-2013 | Disposable Temperature Probes, (sterile), Sensor diameter: 3mm/0.12 in; 25 units per box… GE Healthcare has become aware of a potential safety issue associated with the use of GE General Purpose 9 French Temperature Probe… |
| Oct 16, 2012 | Class II | Z-0330-2013 | Corometrics Qwik Connect Plus Spiral Electrode- Model #7000AA0
A fetal scalp circular… Thirty fetal scalp electrodes were found to have an incomplete package seal and lacked the printed lot number stamped on during sealing. |
| Feb 20, 2012 | Class II | Z-1972-2012 | enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL
100, EFLOW DISPOSABLE… The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1959-2013, this FEI as the recalling firm | Jul 18, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0330-2013, this FEI as the recalling firm | Oct 16, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1972-2012, this FEI as the recalling firm | Feb 20, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 15, 2011 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Dec 1, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing Vital Signs Devices, A GE Healthcare Company as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Vital Signs Devices, A GE Healthcare Company yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Vital Signs Devices, A GE Healthcare Company
- Location
- Totowa, NJ (US)
- Source
- inspections