Topcon Healthcare, Inc. FDA Inspection Record
Topcon Healthcare, Inc. is an FDA-registered facility (FEI 2242863) located in Oakland, NJ, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 14, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Sep 17, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 14, 2025 | Voluntary Action Indicated (VAI) | Quality audits - auditor independence (21 CFR 820.22) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate procedures (21 CFR 820.100(a)) Complaints (21 CFR 820.198(a)) and 3 more |
| Jan 20, 2015 | No Action Indicated (NAI) | No citations listed |
| Jun 16, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 17, 2024 | Class II | Z-0362-2025 | The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in… The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not… |
| Dec 18, 2020 | Class II | Z-0952-2021 | Harmony Referral System (aka Harmony RS)/ Medical Device
Data System - Product Usage… Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused… |
| Mar 6, 2019 | Class II | Z-1138-2019 | IMAGEnet 6 v1.53-
IMAGEnet 6 Ophthalmic Data System is a software program that is… When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is… |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 14, 2025 | 6 x 1 | 6 |
| Class II recall Z-0362-2025, this FEI as the recalling firm | Sep 17, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0952-2021, this FEI as the recalling firm | Dec 18, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1138-2019, this FEI as the recalling firm | Mar 6, 2019 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jun 16, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 14, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 2, 2026 | Newest record Oct 22, 2026 | Current |
Reviewing Topcon Healthcare, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Topcon Healthcare, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Topcon Medical Systems, Inc.
- Location
- Oakland, NJ (US)
- Source