Mti Precision Products FDA Inspection Record

FEI 2246821 · Inspections, Form 483s, warning letters and import refusals

Mti Precision Products is an FDA-registered facility (FEI 2246821) located in Coatesville, PA, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Mar 10, 2023: No Action Indicated (NAI)
Inspections on record
5 (3 OAI, 1 VAI, 1 NAI)
Last OAI inspection
Sep 7, 2016
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jun 6, 2014. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
1, latest May 7, 2012

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 10, 2023No Action Indicated (NAI)No citations listed
Mar 16, 2017Voluntary Action Indicated (VAI)Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
Personnel (21 CFR 820.25(a))
Document review, approval by designated individual (21 CFR 820.40(a))
Sep 7, 2016Official Action Indicated (OAI)Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Documentation (21 CFR 820.80(e))
Complaints (21 CFR 820.198(a))
and 4 more
Nov 6, 2014Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 4 more
Apr 30, 2014Official Action Indicated (OAI)DHR - not or inadequately maintained (21 CFR 820.184)
Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 7 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 7, 2012Class IIZ-2044-2014Lynx TM20 Torquemaster Low Speed Handpiece MTI Dental Products
Sharp edge on the body/housing of the Lynx TM20 TorqueMaster Low Speed Handpiece.

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

6 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIMar 16, 20176 x 0.251.5
Inspection classified OAISep 7, 201610 years or older, so it adds no points0
Inspection classified OAINov 6, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJun 6, 201410 years or older, so it adds no points0
Inspection classified OAIApr 30, 201410 years or older, so it adds no points0
Class II recall Z-2044-2014, this FEI as the recalling firmMay 7, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Mti Precision Products as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Mti Precision Products: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Mti Precision Products → Check this supplier

Facility details

FDA Legal Name
MTI PRECISION PRODUCTS
Company
MTI Precision Products LLC.
Location
Coatesville, PA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.