MP Biomedicals Diagnostics Division FDA Inspection Record

FEI 2419955 · Inspections, Form 483s, warning letters and import refusals

MP Biomedicals Diagnostics Division is an FDA-registered facility (FEI 2419955) located in Solon, OH, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Mar 31, 2026: Voluntary Action Indicated (VAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 2, 2010. Many letters do not state an FEI: search warning letters by company name

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 31, 2026Voluntary Action Indicated (VAI)Product realization risk management processes not documented (ISO 13485:2016 Clause 7.1)
Procedures for purchasing process not documented (ISO 13485:2016 Clause 7.4.1)
Measuring equipment not adequately calibrated or verified (ISO 13485:2016 Clause 7.6(a))
Dec 13, 2017No Action Indicated (NAI)Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b))
DHR content (21 CFR 820.184)
Acceptance status (21 CFR 820.86)
Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2))
and 2 more
Dec 15, 2011Voluntary Action Indicated (VAI)Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
Apr 22, 2010No Action Indicated (NAI)Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
Acceptance records (21 CFR 820.184(d))
Investigation of device failures (21 CFR 820.198(c))
Lack of or inadequate process validation (21 CFR 820.75(a))

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

FDA record score

6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

3 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIMar 31, 20266 x 16
Inspection classified VAIDec 15, 201110 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsAug 2, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing MP Biomedicals Diagnostics Division as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for MP Biomedicals Diagnostics Division: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about MP Biomedicals Diagnostics Division → Check this supplier

Facility details

FDA Legal Name
MP BIOMEDICALS DIAGNOSTICS DIVISION
Company
MP Biomedicals Diagnostics Division
Location
Solon, OH (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.