FUJIFILM Healthcare Americas Corporation FDA Inspection Record

FEI 2431293 · Inspections, Form 483s, warning letters and import refusals

FUJIFILM Healthcare Americas Corporation is an FDA-inspected facility (FEI 2431293) located in Wayne, NJ, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jul 2, 2019: Voluntary Action Indicated (VAI)
Inspections on record
4 (1 OAI, 3 VAI)
Last OAI inspection
Apr 9, 2015
Warning letters
1 listed by FDA for this FEI, latest Aug 12, 2015
Form 483 and inspection documents
1 in FDA's reading room
Recalls as the recalling firm
8, latest Jul 21, 2017
Import refusals
5

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jul 2, 2019Voluntary Action Indicated (VAI)Records of complaint investigation (21 CFR 820.198(e))
Apr 9, 2015Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
Report of risk to health (21 CFR 806.10(a)(1))
and 1 more
Jan 16, 2014Voluntary Action Indicated (VAI)Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Lack of or inadequate procedures (21 CFR 820.100(a))
Complaints (21 CFR 820.198(a))
Dec 13, 2010Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Review and evaluation for investigation (21 CFR 820.198(b))
Investigation of device failures (21 CFR 820.198(c))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 21, 2017Class IIZ-3225-2017FUJINON ED-530XT. This device is intended for the visualization of the duodenum and…
An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device…
Feb 17, 2017Class IIZ-2706-2017Ultrasonic Endoscope Model Number EG-530UT A flexible ultrasonic endoscope intended to…
Fujifilm is recalling multiple endoscopes after a retrospective review.
Dec 23, 2015Class IIZ-1860-2016Endoscope; ED-530XT Operation Manual 'Operation and Preparation' and 'Cleaning… and 4 more products
This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde…
Aug 31, 2015Class IIZ-0418-2016Fujifilm Medical Systems Duodenoscope, Model Numbers ED-200XU, ED-200XT, ED-250XT…
A precautionary measure because the Duodenoscopes have been redesigned to incorporate a closed elevator wire channel that facilitates…
Jun 30, 2015Class IIZ-2778-2015Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5…
The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents…
Jun 30, 2015Class IIZ-2805-2015Fujinon/Fujifilm Processor EPX 2500 Operation Manual. For use in endoscopic procedures. and 1 more product
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500…
Mar 9, 2015Class IIZ-3128-2017Fujifilm Endoscope Model ED-530XT Product Usage: This product is a medical… and 1 more product
Fujifilm is reporting a corrective action due to an FDA inspection.
Jul 31, 2013Class IIZ-0112-2014Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of…
A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely…

Form 483s and inspection documents

  • 483 on FDA.gov
    Inspection Mar 24, 2015 to Apr 9, 2015, issued Apr 20, 2015 (dates printed on the document). FDA listing date Aug 22, 2018. Published by FDA Jan 18, 2024.

FDA record score

4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 16, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

13 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJul 2, 20196 x 0.251.5
Class II recall Z-3225-2017, this FEI as the recalling firmJul 21, 20174 x 0.251
Class II recall Z-2706-2017, this FEI as the recalling firmFeb 17, 20174 x 0.251
Class II recall Z-1860-2016, this FEI as the recalling firmDec 23, 201510 years or older, so it adds no points0
Class II recall Z-0418-2016, this FEI as the recalling firmAug 31, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsAug 12, 201510 years or older, so it adds no points0
Class II recall Z-2778-2015, this FEI as the recalling firmJun 30, 201510 years or older, so it adds no points0
Class II recall Z-2805-2015, this FEI as the recalling firmJun 30, 201510 years or older, so it adds no points0
Inspection classified OAIApr 9, 201510 years or older, so it adds no points0
Class II recall Z-3128-2017, this FEI as the recalling firmMar 9, 201510 years or older, so it adds no points0
Inspection classified VAIJan 16, 201410 years or older, so it adds no points0
Class II recall Z-0112-2014, this FEI as the recalling firmJul 31, 201310 years or older, so it adds no points0
Inspection classified VAIDec 13, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing FUJIFILM Healthcare Americas Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check FUJIFILM Healthcare Americas Corporation yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
FUJIFILM Healthcare Americas Corporation
Location
Wayne, NJ (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.