FUJIFILM Healthcare Americas Corporation FDA Inspection Record
FUJIFILM Healthcare Americas Corporation is an FDA-inspected facility (FEI 2431293) located in Wayne, NJ, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 2, 2019: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (1 OAI, 3 VAI)
- Last OAI inspection
- Apr 9, 2015
- Warning letters
- 1 listed by FDA for this FEI, latest Aug 12, 2015
- Form 483 and inspection documents
- 1 in FDA's reading room
- Recalls as the recalling firm
- 8, latest Jul 21, 2017
- Import refusals
- 5
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 2, 2019 | Voluntary Action Indicated (VAI) | Records of complaint investigation (21 CFR 820.198(e)) |
| Apr 9, 2015 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Report of risk to health (21 CFR 806.10(a)(1)) and 1 more |
| Jan 16, 2014 | Voluntary Action Indicated (VAI) | Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3)) Lack of or inadequate procedures (21 CFR 820.100(a)) Complaints (21 CFR 820.198(a)) |
| Dec 13, 2010 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Review and evaluation for investigation (21 CFR 820.198(b)) Investigation of device failures (21 CFR 820.198(c)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 21, 2017 | Class II | Z-3225-2017 | FUJINON ED-530XT.
This device is intended for the visualization of the duodenum and… An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device… |
| Feb 17, 2017 | Class II | Z-2706-2017 | Ultrasonic Endoscope
Model Number EG-530UT
A flexible ultrasonic endoscope intended to… Fujifilm is recalling multiple endoscopes after a retrospective review. |
| Dec 23, 2015 | Class II | Z-1860-2016 | Endoscope; ED-530XT Operation Manual
'Operation and Preparation' and 'Cleaning… and 4 more products This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde… |
| Aug 31, 2015 | Class II | Z-0418-2016 | Fujifilm Medical Systems Duodenoscope, Model Numbers ED-200XU, ED-200XT, ED-250XT… A precautionary measure because the Duodenoscopes have been redesigned to incorporate a closed elevator wire channel that facilitates… |
| Jun 30, 2015 | Class II | Z-2778-2015 | Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5… The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents… |
| Jun 30, 2015 | Class II | Z-2805-2015 | Fujinon/Fujifilm Processor EPX 2500 Operation Manual. For use in endoscopic procedures. and 1 more product The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500… |
| Mar 9, 2015 | Class II | Z-3128-2017 | Fujifilm Endoscope
Model ED-530XT
Product Usage:
This product is a medical… and 1 more product Fujifilm is reporting a corrective action due to an FDA inspection. |
| Jul 31, 2013 | Class II | Z-0112-2014 | Fujinon/Fujifilm Ultrasonic Endoscope
Intended to provide ultrasonic images of… A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely… |
Form 483s and inspection documents
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
13 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jul 2, 2019 | 6 x 0.25 | 1.5 |
| Class II recall Z-3225-2017, this FEI as the recalling firm | Jul 21, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2706-2017, this FEI as the recalling firm | Feb 17, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1860-2016, this FEI as the recalling firm | Dec 23, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0418-2016, this FEI as the recalling firm | Aug 31, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 12, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2778-2015, this FEI as the recalling firm | Jun 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2805-2015, this FEI as the recalling firm | Jun 30, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 9, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-3128-2017, this FEI as the recalling firm | Mar 9, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 16, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0112-2014, this FEI as the recalling firm | Jul 31, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 13, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing FUJIFILM Healthcare Americas Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check FUJIFILM Healthcare Americas Corporation yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- FUJIFILM Healthcare Americas Corporation
- Location
- Wayne, NJ (US)
- Source
- inspections