MIRION TECHNOLOGIES (CAPINTEC), INC FDA Inspection Record
MIRION TECHNOLOGIES (CAPINTEC), INC is an FDA-registered facility (FEI 2518443) located in Florham Park, NJ, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 25, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Jul 20, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 25, 2016 | Voluntary Action Indicated (VAI) | No citations listed |
| Jan 23, 2013 | No Action Indicated (NAI) | No citations listed |
| Dec 8, 2010 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Rationale documented for no investigation (21 CFR 820.198(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 20, 2026 | Class II | Z-3193-2026 | Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154; Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is… |
| Apr 22, 2024 | Class II | Z-1892-2024 | Captus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff… and 2 more products Complaints of unexpected detachment of the collimator have been reported. |
| Nov 27, 2023 | Class II | Z-1069-2024 | Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine… Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel… |
| Mar 3, 2020 | Class II | Z-1564-2020 | Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product… It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator… |
| Jul 24, 2012 | Class II | Z-2312-2012 | Capintec Captus 3000 Thyroid Uptake System
Nuclear uptake probe intended to be used by… Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter… |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-3193-2026, this FEI as the recalling firm | Jul 20, 2026 | 4 x 1 | 4 |
| Class II recall Z-1892-2024, this FEI as the recalling firm | Apr 22, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1069-2024, this FEI as the recalling firm | Nov 27, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1564-2020, this FEI as the recalling firm | Mar 3, 2020 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jan 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2312-2012, this FEI as the recalling firm | Jul 24, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 8, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing MIRION TECHNOLOGIES (CAPINTEC), INC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check MIRION TECHNOLOGIES (CAPINTEC), INC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Mirion Technologies (Capintec), Inc.
- Location
- Florham Park, NJ (US)
- Source
- device