Varian Medical Systems, Inc FDA Inspection Record

FEI 2916710 · Inspections, Form 483s, warning letters and import refusals

Varian Medical Systems, Inc is an FDA-registered facility (FEI 2916710) located in PALO ALTO, CA, US. RegDataLab tracks 6 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 15, 2016: No Action Indicated (NAI)
Inspections on record
6 (6 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
22, latest Aug 12, 2019
Import refusals
1

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 15, 2016No Action Indicated (NAI)No citations listed
Aug 10, 2016No Action Indicated (NAI)No citations listed
Jul 10, 2015No Action Indicated (NAI)No citations listed
Jun 6, 2014No Action Indicated (NAI)No citations listed
Apr 30, 2013No Action Indicated (NAI)No citations listed
Oct 18, 2010No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 22 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 12, 2019Class IIZ-1899-2020Varian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High…
After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move…
Dec 26, 2018Class IIZ-0771-2019Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm…
There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm…
Dec 2, 2016Class IIZ-0990-2017Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation…
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved…
Sep 13, 2016Class IIZ-0077-2017Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits…
Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on…
Jun 23, 2016Class IIZ-2149-2016Optical Surface Monitoring System{OSMS, Varian Private Label} This is used on Varian's… and 1 more product
Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation…
Jan 11, 2016Class IIZ-0805-2016Eclipse Treatment Planning System versions 11, 13.0, 13.5 and 13.6; Model number…
When using PBC 11.0.31 to calculate the dose for a conventional arc field with more than 100 segments for Eclipse versions 11.0, 13.0, 13.5…
Sep 28, 2015Class IIZ-0170-2016ARIA Radiation Oncology, versions 10, 11 , 13.0 and 13.5 with Clinical Assessment License…
An anomaly was identified with the ARIA for Radiation Oncology software with a Clinical Assessment license. Admin Instructions may not print…
Aug 18, 2014Class IIZ-2654-20144D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided…
Due to user error, the 4DITC can improperly allow users to clear a safety interlock imposed by the 4DITC on behalf of Varian's Bar Code…
Jun 17, 2014Class IIZ-2567-2014C-Series High Energy Linear Accelerator - Models: Novalis Tx, Trilogy, Trilogy Tx, Clinac…
Trend in reports of unexpected decrease in beam output in C-series High Energy Linear Accelerators for 6MV photon treatment mode.
Apr 30, 2014Class IIZ-0526-2015Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for…
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground…

FDA record score

3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

22 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1899-2020, this FEI as the recalling firmAug 12, 20194 x 0.251
Class II recall Z-0771-2019, this FEI as the recalling firmDec 26, 20184 x 0.251
Class II recall Z-0990-2017, this FEI as the recalling firmDec 2, 20164 x 0.251
Class II recall Z-0077-2017, this FEI as the recalling firmSep 13, 201610 years or older, so it adds no points0
Class II recall Z-2149-2016, this FEI as the recalling firmJun 23, 201610 years or older, so it adds no points0
Class II recall Z-0805-2016, this FEI as the recalling firmJan 11, 201610 years or older, so it adds no points0
Class II recall Z-0170-2016, this FEI as the recalling firmSep 28, 201510 years or older, so it adds no points0
Class II recall Z-2654-2014, this FEI as the recalling firmAug 18, 201410 years or older, so it adds no points0
Class II recall Z-2567-2014, this FEI as the recalling firmJun 17, 201410 years or older, so it adds no points0
Class II recall Z-0526-2015, this FEI as the recalling firmApr 30, 201410 years or older, so it adds no points0
Class II recall Z-1502-2014, this FEI as the recalling firmMar 20, 201410 years or older, so it adds no points0
Class II recall Z-0552-2014, this FEI as the recalling firmNov 18, 201310 years or older, so it adds no points0
Class II recall Z-0439-2014, this FEI as the recalling firmOct 24, 201310 years or older, so it adds no points0
Class II recall Z-1163-2013, this FEI as the recalling firmMar 26, 201310 years or older, so it adds no points0
Class II recall Z-0867-2013, this FEI as the recalling firmJan 28, 201310 years or older, so it adds no points0
Class II recall Z-0531-2013, this FEI as the recalling firmNov 5, 201210 years or older, so it adds no points0
Class II recall Z-0187-2013, this FEI as the recalling firmSep 24, 201210 years or older, so it adds no points0
Class II recall Z-2462-2012, this FEI as the recalling firmAug 17, 201210 years or older, so it adds no points0
Class II recall Z-2383-2012, this FEI as the recalling firmAug 14, 201210 years or older, so it adds no points0
Class II recall Z-2153-2012, this FEI as the recalling firmJul 2, 201210 years or older, so it adds no points0
Class II recall Z-2041-2012, this FEI as the recalling firmJun 18, 201210 years or older, so it adds no points0
Class II recall Z-1325-2013, this FEI as the recalling firmMay 31, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Varian Medical Systems, Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Varian Medical Systems, Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Varian Medical Systems, Inc.
Location
PALO ALTO, CA (US)
Source
device