Hamilton Medical, Inc. FDA Inspection Record

FEI 2937708 · Inspections, Form 483s, warning letters and import refusals

Hamilton Medical, Inc. is an FDA-inspected facility (FEI 2937708) located in Reno, NV, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
May 17, 2019: No Action Indicated (NAI)
Inspections on record
4 (4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
10, latest Jul 26, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
May 17, 2019No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Lack of or inadequate procedures (21 CFR 820.100(a))
Management review - Lack of or inadequate procedures (21 CFR 820.20(c))
Oct 2, 2015No Action Indicated (NAI)No citations listed
Jun 6, 2013No Action Indicated (NAI)No citations listed
Apr 5, 2012No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 26, 2023Class IZ-2525-2023Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator and 5 more products
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit…
Jun 15, 2023Class IZ-2139-2023HAMILTON C1 Ventilator REF 161001 1610010 Is intended to provide positive pressure… and 3 more products
Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior…
Jun 20, 2017Class IIZ-0335-2018Intellicuff Standalone (PN 951001) Used to continuously measure and automatically…
Issue related to the performance of the motor in recalled product. During use, motor may cease to function.
Sep 26, 2016Class IIZ-0666-2017Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to…
Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.
Mar 9, 2016Class IIZ-1624-2016Hamilton G5 with software version between v2.41, v2.42, v2.50. Catalog number: 1590001…
After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern…
Jun 1, 2015Class IIZ-1963-2015Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177. The infant… and 2 more products
An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control…
Mar 24, 2015Class IIZ-2057-2015Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001…
Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and…
Apr 22, 2014Class IZ-0151-2016HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology…
Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the…
Sep 26, 2013Class IIZ-0170-2014HAMILTON-C2 Ventilator: Manufactured by Hamilton Medical AG CH-7402 Bonaduz…
Manufacturer defined limits for improperly functioning loudspeakers of the HAMILTON C2 alarm system were exceeded.
Dec 19, 2012Class IZ-0741-2013HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1…
Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung…

FDA record score

19 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-2525-2023, this FEI as the recalling firmJul 26, 202315 x 0.69
Class I recall Z-2139-2023, this FEI as the recalling firmJun 15, 202315 x 0.69
Class II recall Z-0335-2018, this FEI as the recalling firmJun 20, 20174 x 0.251
Class II recall Z-0666-2017, this FEI as the recalling firmSep 26, 201610 years or older, so it adds no points0
Class II recall Z-1624-2016, this FEI as the recalling firmMar 9, 201610 years or older, so it adds no points0
Class II recall Z-1963-2015, this FEI as the recalling firmJun 1, 201510 years or older, so it adds no points0
Class II recall Z-2057-2015, this FEI as the recalling firmMar 24, 201510 years or older, so it adds no points0
Class I recall Z-0151-2016, this FEI as the recalling firmApr 22, 201410 years or older, so it adds no points0
Class II recall Z-0170-2014, this FEI as the recalling firmSep 26, 201310 years or older, so it adds no points0
Class I recall Z-0741-2013, this FEI as the recalling firmDec 19, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Hamilton Medical, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Hamilton Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Hamilton Medical, Inc. → Check this supplier

Facility details

Company
Hamilton Medical, Inc.
Location
Reno, NV (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.