Hamilton Medical, Inc. FDA Inspection Record
Hamilton Medical, Inc. is an FDA-inspected facility (FEI 2937708) located in Reno, NV, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 17, 2019: No Action Indicated (NAI)
- Inspections on record
- 4 (4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 10, latest Jul 26, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 17, 2019 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Lack of or inadequate procedures (21 CFR 820.100(a)) Management review - Lack of or inadequate procedures (21 CFR 820.20(c)) |
| Oct 2, 2015 | No Action Indicated (NAI) | No citations listed |
| Jun 6, 2013 | No Action Indicated (NAI) | No citations listed |
| Apr 5, 2012 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 26, 2023 | Class I | Z-2525-2023 | Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator and 5 more products Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit… |
| Jun 15, 2023 | Class I | Z-2139-2023 | HAMILTON C1 Ventilator REF
161001
1610010
Is intended to provide positive pressure… and 3 more products Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior… |
| Jun 20, 2017 | Class II | Z-0335-2018 | Intellicuff Standalone (PN 951001)
Used to continuously measure and automatically… Issue related to the performance of the motor in recalled product. During use, motor may cease to function. |
| Sep 26, 2016 | Class II | Z-0666-2017 | Hamilton-MR1 Ventilator: Catalog# 161010
The Hamilton MR1 Ventilator is intended to… Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation. |
| Mar 9, 2016 | Class II | Z-1624-2016 | Hamilton G5 with software version between v2.41, v2.42, v2.50.
Catalog number: 1590001… After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern… |
| Jun 1, 2015 | Class II | Z-1963-2015 | Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177. The infant… and 2 more products An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control… |
| Mar 24, 2015 | Class II | Z-2057-2015 | Hamilton-G5 Ventilators with software versions V2.40/2.41
Catalog numbers for G5: 159001… Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and… |
| Apr 22, 2014 | Class I | Z-0151-2016 | HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31.
Anesthesiology… Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the… |
| Sep 26, 2013 | Class II | Z-0170-2014 | HAMILTON-C2 Ventilator:
Manufactured by Hamilton Medical AG
CH-7402 Bonaduz… Manufacturer defined limits for improperly functioning loudspeakers of the HAMILTON C2 alarm system were exceeded. |
| Dec 19, 2012 | Class I | Z-0741-2013 | HAMILTON-T1 Ventilator with software versions 1.1.2 or lower.
The HAMILTON T-1… Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung… |
FDA record score
19 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-2525-2023, this FEI as the recalling firm | Jul 26, 2023 | 15 x 0.6 | 9 |
| Class I recall Z-2139-2023, this FEI as the recalling firm | Jun 15, 2023 | 15 x 0.6 | 9 |
| Class II recall Z-0335-2018, this FEI as the recalling firm | Jun 20, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0666-2017, this FEI as the recalling firm | Sep 26, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1624-2016, this FEI as the recalling firm | Mar 9, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1963-2015, this FEI as the recalling firm | Jun 1, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2057-2015, this FEI as the recalling firm | Mar 24, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0151-2016, this FEI as the recalling firm | Apr 22, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0170-2014, this FEI as the recalling firm | Sep 26, 2013 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0741-2013, this FEI as the recalling firm | Dec 19, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Hamilton Medical, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Hamilton Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Hamilton Medical, Inc. → Check this supplier
Facility details
- Company
- Hamilton Medical, Inc.
- Location
- Reno, NV (US)
- Source
- inspections