Volcano Corporation FDA Inspection Record

FEI 2939520 · Inspections, Form 483s, warning letters and import refusals

Volcano Corporation is an FDA-registered facility (FEI 2939520) located in Rancho Cordova, CA, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jan 8, 2025: No Action Indicated (NAI)
Inspections on record
5 (2 VAI, 3 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
13, latest Jan 19, 2022
Import refusals
4

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jan 8, 2025No Action Indicated (NAI)No citations listed
May 28, 2015No Action Indicated (NAI)No citations listed
Oct 30, 2012Voluntary Action Indicated (VAI)DHR content (21 CFR 820.184)
Dec 22, 2011No Action Indicated (NAI)No citations listed
Sep 28, 2009Voluntary Action Indicated (VAI)No citations listed

Recalls

Showing the 10 most recent of 13 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 19, 2022Class IIZ-0885-2022SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x
If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect…
Nov 7, 2018Class IIZ-0674-2020Philips Volcano FFR software used in the following systems connected to the McKesson…
Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems…
Oct 17, 2018Class IIZ-0726-2019Visions PV .035 Digital IVUS Catheter, Catalog Number/Part Number… and 1 more product
Visions PV .035 Intravascular Ultrasound (IVUS) Catheters (catalog number 88901 adnd81234) due to the potential inability to pass the…
Mar 22, 2018Class IIZ-1891-2018Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the…
The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.
Nov 3, 2017Class IIZ-0189-2018Volcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF…
On certain Impacted Systems, an unexpected Microsoft Windows Security dialog may appear during use, and the users response to the dialog may…
Mar 29, 2017Class IIZ-1775-2017Refinity Rotational Intravascular Ultrasound (IVUS) Catheter; Product number… and 1 more product
High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in…
Apr 25, 2016Class IIZ-1813-2016Volcano s5, s5i, CORE and CORE Mobile systems with software version 3.4 and v3.4 software…
Volcano has become aware of an incompatibility issue between Impacted Systems and hospital network scans. In specific circumstances, an…
Feb 5, 2016Class IIZ-1041-2016Volcano R100 Sterile Equipment Cover; Sterile; General Hospital: These Sterile… and 1 more product
Volcano Corporation has identified that certain lot numbers of Volcano R-100 Sterile Equipment Covers may have the incorrect expiry date on…
Jul 22, 2015Class IIZ-2718-2015Volcano Pioneer Plus Re-Entry Catheter; Product Code PPLUS20, Catheter for Crossing Total…
Due to a manufacturing defect, there is the possibility that a small wire could extend through the catheter shaft of the Pioneer Plus…
Jun 22, 2015Class IIZ-2131-2015Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4… and 5 more products
A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

FDA record score

7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 28, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

15 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0885-2022, this FEI as the recalling firmJan 19, 20224 x 0.62.4
Class II recall Z-0674-2020, this FEI as the recalling firmNov 7, 20184 x 0.251
Class II recall Z-0726-2019, this FEI as the recalling firmOct 17, 20184 x 0.251
Class II recall Z-1891-2018, this FEI as the recalling firmMar 22, 20184 x 0.251
Class II recall Z-0189-2018, this FEI as the recalling firmNov 3, 20174 x 0.251
Class II recall Z-1775-2017, this FEI as the recalling firmMar 29, 20174 x 0.251
Class II recall Z-1813-2016, this FEI as the recalling firmApr 25, 201610 years or older, so it adds no points0
Class II recall Z-1041-2016, this FEI as the recalling firmFeb 5, 201610 years or older, so it adds no points0
Class II recall Z-2718-2015, this FEI as the recalling firmJul 22, 201510 years or older, so it adds no points0
Class II recall Z-2131-2015, this FEI as the recalling firmJun 22, 201510 years or older, so it adds no points0
Class II recall Z-1844-2015, this FEI as the recalling firmJun 4, 201510 years or older, so it adds no points0
Class II recall Z-2636-2014, this FEI as the recalling firmAug 13, 201410 years or older, so it adds no points0
Inspection classified VAIOct 30, 201210 years or older, so it adds no points0
Class II recall Z-2345-2012, this FEI as the recalling firmAug 24, 201210 years or older, so it adds no points0
Inspection classified VAISep 28, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 28, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 28, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 28, 2026Newest record Nov 19, 2026Current
Warning letter textNov 28, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 28, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Volcano Corporation as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Volcano Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Volcano Corporation → Check this supplier

Facility details

FDA Legal Name
VOLCANO CORPORATION
Company
Philips Image Guided Therapy Corporation
Location
Rancho Cordova, CA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.