Volcano Corporation FDA Inspection Record
Volcano Corporation is an FDA-registered facility (FEI 2939520) located in Rancho Cordova, CA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 8, 2025: No Action Indicated (NAI)
- Inspections on record
- 5 (2 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest Jan 19, 2022
- Import refusals
- 4
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 8, 2025 | No Action Indicated (NAI) | No citations listed |
| May 28, 2015 | No Action Indicated (NAI) | No citations listed |
| Oct 30, 2012 | Voluntary Action Indicated (VAI) | DHR content (21 CFR 820.184) |
| Dec 22, 2011 | No Action Indicated (NAI) | No citations listed |
| Sep 28, 2009 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 19, 2022 | Class II | Z-0885-2022 | SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect… |
| Nov 7, 2018 | Class II | Z-0674-2020 | Philips Volcano FFR software used in the following systems connected to the McKesson… Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems… |
| Oct 17, 2018 | Class II | Z-0726-2019 | Visions PV .035 Digital IVUS Catheter, Catalog Number/Part Number… and 1 more product Visions PV .035 Intravascular Ultrasound (IVUS) Catheters (catalog number 88901 adnd81234) due to the potential inability to pass the… |
| Mar 22, 2018 | Class II | Z-1891-2018 | Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only.
Used for the… The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image. |
| Nov 3, 2017 | Class II | Z-0189-2018 | Volcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF… On certain Impacted Systems, an unexpected Microsoft Windows Security dialog may appear during use, and the users response to the dialog may… |
| Mar 29, 2017 | Class II | Z-1775-2017 | Refinity Rotational Intravascular Ultrasound (IVUS) Catheter;
Product number… and 1 more product High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in… |
| Apr 25, 2016 | Class II | Z-1813-2016 | Volcano s5, s5i, CORE and CORE Mobile systems with software version 3.4 and v3.4 software… Volcano has become aware of an incompatibility issue between Impacted Systems and hospital network scans. In specific circumstances, an… |
| Feb 5, 2016 | Class II | Z-1041-2016 | Volcano R100 Sterile Equipment Cover; Sterile;
General Hospital: These Sterile… and 1 more product Volcano Corporation has identified that certain lot numbers of Volcano R-100 Sterile Equipment Covers may have the incorrect expiry date on… |
| Jul 22, 2015 | Class II | Z-2718-2015 | Volcano Pioneer Plus Re-Entry Catheter;
Product Code PPLUS20, Catheter for Crossing Total… Due to a manufacturing defect, there is the possibility that a small wire could extend through the catheter shaft of the Pioneer Plus… |
| Jun 22, 2015 | Class II | Z-2131-2015 | Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4… and 5 more products A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances. |
FDA record score
7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
15 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0885-2022, this FEI as the recalling firm | Jan 19, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0674-2020, this FEI as the recalling firm | Nov 7, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0726-2019, this FEI as the recalling firm | Oct 17, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1891-2018, this FEI as the recalling firm | Mar 22, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0189-2018, this FEI as the recalling firm | Nov 3, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1775-2017, this FEI as the recalling firm | Mar 29, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1813-2016, this FEI as the recalling firm | Apr 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1041-2016, this FEI as the recalling firm | Feb 5, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2718-2015, this FEI as the recalling firm | Jul 22, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2131-2015, this FEI as the recalling firm | Jun 22, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1844-2015, this FEI as the recalling firm | Jun 4, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2636-2014, this FEI as the recalling firm | Aug 13, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 30, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2345-2012, this FEI as the recalling firm | Aug 24, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 28, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 28, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Volcano Corporation as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Volcano Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- FDA Legal Name
- VOLCANO CORPORATION
- Company
- Philips Image Guided Therapy Corporation
- Location
- Rancho Cordova, CA (US)
- Source
- device