BECKMAN COULTER, INC. FDA Inspection Record
BECKMAN COULTER, INC. is an FDA-registered facility (FEI 3000203293) located in Miami, FL, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Oct 28, 2021: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (1 OAI, 2 VAI, 1 NAI)
- Last OAI inspection
- Jun 2, 2011
- Warning letters
- 1 listed by FDA for this FEI, latest Nov 1, 2011
- Recalls as the recalling firm
- 13, latest Aug 12, 2026
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 28, 2021 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Oct 16, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Servicing - Lack of or inadequate procedures (21 CFR 820.200(a)) Lack of Written MDR Procedures (21 CFR 803.17) |
| Sep 11, 2013 | No Action Indicated (NAI) | No citations listed |
| Jun 2, 2011 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Procedures not adequately established or maintained (21 CFR 820.40) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) and 3 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 12, 2026 | Class II | Z-0160-2027 | DxU 840m Iris, REF: C76947;
DxU 850m Iris, REF: C49513 and 1 more product The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid… |
| Dec 23, 2025 | Class II | Z-1235-2026 | CellMek SPS Sample Preparation System, REF: C44603, with software Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close… |
| Nov 10, 2025 | Class II | Z-1221-2026 | COULTER DxH Diluent, REF 628017
COULTER DxH ECO Diluent REF C67250
For use as an… Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers. |
| Sep 8, 2025 | Class II | Z-0412-2026 | CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior… Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect… |
| Jul 16, 2024 | Class III | Z-0093-2025 | Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and… Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure… |
| Dec 19, 2023 | Class II | Z-0784-2024 | Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947 and 3 more products Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust… |
| May 31, 2023 | Class II | Z-2138-2023 | COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications. |
| Apr 5, 2023 | Class II | Z-1596-2023 | iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and… and 3 more products Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not… |
| Nov 18, 2022 | Class II | Z-0777-2023 | BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic
use only. Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive… |
| Feb 16, 2022 | Class III | Z-0984-2022 | Beckman Coulter IRISpec CA/CB/CC control, Catalog #800-7211, containing 3 x 100 mL bottles… IRISpec CA failures for glucose. |
FDA record score
32 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
15 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0160-2027, this FEI as the recalling firm | Aug 12, 2026 | 4 x 1 | 4 |
| Class II recall Z-1235-2026, this FEI as the recalling firm | Dec 23, 2025 | 4 x 1 | 4 |
| Class II recall Z-1221-2026, this FEI as the recalling firm | Nov 10, 2025 | 4 x 1 | 4 |
| Class II recall Z-0412-2026, this FEI as the recalling firm | Sep 8, 2025 | 4 x 1 | 4 |
| Class II recall Z-0784-2024, this FEI as the recalling firm | Dec 19, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2138-2023, this FEI as the recalling firm | May 31, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1596-2023, this FEI as the recalling firm | Apr 5, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0777-2023, this FEI as the recalling firm | Nov 18, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Oct 28, 2021 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Oct 16, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2241-2021, this FEI as the recalling firm | Jun 23, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1535-2020, this FEI as the recalling firm | Feb 17, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0516-2020, this FEI as the recalling firm | Oct 7, 2019 | 4 x 0.25 | 1 |
| Warning letter on FDA's dashboard, which does not show close-outs | Nov 1, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jun 2, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 14, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 2, 2026 | Newest record Oct 22, 2026 | Current |
Reviewing BECKMAN COULTER, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check BECKMAN COULTER, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Beckman Coulter, Inc.
- Location
- Miami, FL (US)
- Source
- device