WalkMed Technologies, LLC FDA Inspection Record

FEI 3000204189 · Inspections, Form 483s, warning letters and import refusals

WalkMed Technologies, LLC is an FDA-registered facility (FEI 3000204189) located in Aurora, CO, US. RegDataLab tracks 6 FDA inspections and 2 compliance actions at this facility.

What FDA found at this site

Latest inspection
Sep 13, 2024: Voluntary Action Indicated (VAI)
Inspections on record
6 (1 OAI, 3 VAI, 2 NAI)
Last OAI inspection
Jun 11, 2015
Warning letters
2 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Nov 2, 2015. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Sep 27, 2017

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 13, 2024Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Lack of System for Event Evaluations (21 CFR 803.17(a)(1))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Oct 21, 2021No Action Indicated (NAI)Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Jun 26, 2018Voluntary Action Indicated (VAI)Records of complaint investigation (21 CFR 820.198(e))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
Jun 11, 2015Official Action Indicated (OAI)Lack of or inadequate procedures for distribution (21 CFR 820.160(a))
Design input - documentation (21 CFR 820.30(c))
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 6 more
Apr 15, 2014No Action Indicated (NAI)Procedures not adequately established or maintained (21 CFR 820.40)
Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
Remedial action (21 CFR 820.72(b))
and 4 more
Sep 16, 2010Voluntary Action Indicated (VAI)Design control - no procedures (21 CFR 820.30(a))
Investigation of device failures (21 CFR 820.198(c))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Report of risk to health (21 CFR 806.10(a)(1))
and 1 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Sep 27, 2017Class IIZ-0090-2018Reservoir Bag for WalkMed Infusion Pump Part Number 204820 Catalog Number IPR-86/10 65 mL…
There have been reported incidents of fluid leaks at the luer connection. Leaks have been confirmed to originate from, or near, the…
Jun 14, 2016Class IIZ-2219-2016The pump (with cushioning foam inserts) is packaged in a single pump box. Four pump boxes… and 1 more product
WalkMed Infusion, LLC Announces a Voluntary Field Action of the Triton Infusion Pump (model 300000) and Triton FP Infusion Pump (model…
Jun 14, 2016Class IIZ-2221-2016Spike Set, 12" Microbore SPM-12. Packaged in a Tyvek pouch, with 50 pouches per… and 14 more products
WalkMed Infusion, LLC Announces a Voluntary Field Action of Multiple Triton Administration Sets Due to the Potential Presence of Particulate…
Oct 1, 2015Class IIZ-0242-2016Triton Infusion Pump (model 300000) and Triton fp Infusion Pump (model 400000)…
The device might fail to detect air in line while infusing the medication Venofer. This may occur when a drop of Venofer gets stuck in the…
Feb 17, 2014Class IIZ-1122-2014Triton FP Infusion Pump, Part Number 400000, packaged in boxes. The Triton Volumetric…
WalkMed Infusion is initiaing a recall on their Triton FP Infusion Pump, Part Number 400000 due to the label on the right side of the pump…

FDA record score

6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

11 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAISep 13, 20246 x 0.63.6
Inspection classified VAIJun 26, 20186 x 0.251.5
Class II recall Z-0090-2018, this FEI as the recalling firmSep 27, 20174 x 0.251
Class II recall Z-2219-2016, this FEI as the recalling firmJun 14, 201610 years or older, so it adds no points0
Class II recall Z-2221-2016, this FEI as the recalling firmJun 14, 201610 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsNov 2, 201510 years or older, so it adds no points0
Class II recall Z-0242-2016, this FEI as the recalling firmOct 1, 201510 years or older, so it adds no points0
Inspection classified OAIJun 11, 201510 years or older, so it adds no points0
Class II recall Z-1122-2014, this FEI as the recalling firmFeb 17, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsNov 22, 201010 years or older, so it adds no points0
Inspection classified VAISep 16, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing WalkMed Technologies, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check WalkMed Technologies, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
WalkMed Technologies, LLC
Location
Aurora, CO (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.