Beekley Corporation FDA Inspection Record
Beekley Corporation is an FDA-registered facility (FEI 3000204334) located in Bristol, CT, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 9, 2025: No Action Indicated (NAI)
- Inspections on record
- 5 (2 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 listed by FDA for this FEI, latest Jun 11, 2012
- Recalls as the recalling firm
- 5, latest Jun 13, 2018
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 9, 2025 | No Action Indicated (NAI) | No citations listed |
| Sep 19, 2018 | No Action Indicated (NAI) | Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Apr 28, 2015 | Voluntary Action Indicated (VAI) | Design validation - simulated testing (21 CFR 820.30(g)) |
| Nov 16, 2012 | Voluntary Action Indicated (VAI) | No citations listed |
| May 3, 2012 | No Action Indicated (NAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Investigation of device failures (21 CFR 820.198(c)) Design validation - simulated testing (21 CFR 820.30(g)) and 4 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 13, 2018 | Class II | Z-2830-2018 | O-SPOT REF 791, 791S
Adhere to skin for marking and identifying moles in mammography. and 2 more products There is a potential for imaging artifact to mimic calcifications on patient's mammograms |
| May 8, 2017 | Class III | Z-2309-2017 | Beekley Medical 0-SPOT Mammography Skin Markers
REF 652 Mammography Skin Marker 0 -SPOT (REF 652) product was mis-labeled with A-SPOT (REF 651) labels |
| Jun 19, 2013 | Class III | Z-1787-2014 | Beekley Spots Light Image O-Spots for Mammography
Catalog Number: 252
Product Usage… Light Image O-Spots Dispenser mislabeled as Soft n' Stretchy Light Image S-Spots |
| Jul 9, 2012 | Class II | Z-2109-2012 | Beekley S-Spots Floral Skin Markers -Order Number: 750
Product Usage:
Mammography Skin… and 3 more products Digital Skin Mammography Marker separating from the backing of the spot material |
| Jul 2, 2012 | Class II | Z-2080-2012 | Bella Blanket Protective Coverlet, for Mammography, 18cm x 25cm or 24cm x 30cm, Product… Imaging artifact can occur on mammogram result |
FDA record score
1 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2830-2018, this FEI as the recalling firm | Jun 13, 2018 | 4 x 0.25 | 1 |
| Inspection classified VAI | Apr 28, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 16, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2109-2012, this FEI as the recalling firm | Jul 9, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2080-2012, this FEI as the recalling firm | Jul 2, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jun 11, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 14, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 2, 2026 | Newest record Oct 22, 2026 | Current |
Reviewing Beekley Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Beekley Corporation yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Beekley Corporation
- Location
- Bristol, CT (US)
- Source
- device