INTL. MEDSURG CONNECTION, INC. FDA Inspection Record
INTL. MEDSURG CONNECTION, INC. is an FDA-registered facility (FEI 3000210101) located in Schaumburg, IL, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Oct 17, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 7 (2 OAI, 2 VAI, 3 NAI)
- Last OAI inspection
- Nov 22, 2017
- Warning letters
- 1 listed by FDA for this FEI, latest May 8, 2017
- Recalls as the recalling firm
- 1, latest Jun 23, 2017
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 17, 2025 | Voluntary Action Indicated (VAI) | Lack of System for Event Evaluations (21 CFR 803.17(a)(1)) |
| Aug 30, 2023 | No Action Indicated (NAI) | No citations listed |
| Sep 17, 2018 | No Action Indicated (NAI) | No citations listed |
| Nov 22, 2017 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b)) |
| Sep 16, 2016 | Official Action Indicated (OAI) | Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Lack of or inadequate procedures (21 CFR 820.100(a)) Statistical techniques - Lack of or inadequate procedures (21 CFR 820.250(a)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) and 1 more |
| Dec 12, 2012 | Voluntary Action Indicated (VAI) | Design control - no procedures (21 CFR 820.30(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| May 28, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 23, 2017 | Class II | Z-2804-2017 | Syringe Bulb, 60 CC 50 pcs per case.
The products are used to inject fluids into, or… and 58 more products Insufficient documentation to support product sterility claims. |
FDA record score
22 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified OAI | Nov 22, 2017 | 30 x 0.25 | 7.5 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 8, 2017 | 30 x 0.25 | 7.5 |
| Inspection classified VAI | Oct 17, 2025 | 6 x 1 | 6 |
| Class II recall Z-2804-2017, this FEI as the recalling firm | Jun 23, 2017 | 4 x 0.25 | 1 |
| Inspection classified OAI | Sep 16, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 12, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |